QIAcuity User Manual 06/2022
265
Ensure that no liquid runs down the touchscreen. Liquid may be drawn through the dust
protection sealing by capillary forces and cause malfunction of the display. To clean the
touchscreen, moisten a soft lint-free cloth with water, ethanol, or a mild detergent and carefully
wipe the display. Wipe dry with a paper towel.
CAUTION
Damage to the instrument
[C10]
Do not use bleach, solvents, or reagents containing acids, alkalis, or
abrasives to clean QIAcuity.
CAUTION
Damage to the instrument
[C11]
Do not use spray bottles containing alcohol or disinfectant to clean surfaces
of the QIAcuity. Take special care while cleaning the extended drawer that
no liquid is spilled into the inside of the instrument.
WARNING
Risk of fire
[W24]
Do not allow cleaning fluid or decontamination agents to come into contact
with the electrical parts of the QIAcuity. Take special care while cleaning the
extended drawer that no liquid is spilled into the inside of the instrument.
WARNING
Risk of electric shock
[W24]
Do not open any panels on the QIAcuity.
Risk of personal injury and material damage
Only perform maintenance that is specifically described in this user manual.
WARNING
Hazardous chemicals and infectious agents
[W24]
The plates may contain hazardous material and must be disposed of
properly. Refer to your local safety regulations for proper disposal
procedures.
Содержание QIAcuity 911000
Страница 2: ...QIAcuity User Manual 06 2022 2...
Страница 68: ...QIAcuity User Manual 06 2022 68...
Страница 126: ...QIAcuity User Manual 06 2022 126 Pressing the Users list button will bring you to the list of all users and their status...
Страница 178: ...QIAcuity User Manual 06 2022 178 Raw image of a well showing dust particles marked with red circles...
Страница 207: ...QIAcuity User Manual 06 2022 207 To open an individual plot click on the appropriate header of the 1D Scatterplot...
Страница 240: ...QIAcuity User Manual 06 2022 240...