QIAcuity User Manual 06/2022
209
Changing the thresholds for all selected wells (Common threshold)
The 1D Scatterplot view provides an option to set a common threshold value for all associated wells
using the Common threshold input field. Click the
Recalculate
button to trigger the re-analysis of
data
Use the
Auto-threshold
button to re-set the threshold to the values, which is calculated by the analysis
algorithm for each well. Click the
Recalculate
button to trigger the re-analysis of data.
If you change thresholds and wants to leave the 1D Scatterplot view without triggering the
Recalculate
button a propriate warning message windows opens. You can continue ignoring your
changes or stay in the 1D Scatterplot view.
The Common threshold input field, auto-threshold icon and recalculation button are only shown
when the mouse is in the range of the diagram chart region.
The 1D Scatterplot view with Common threshold.
To add any of the 1D Scatterplot views to the report, click
Add to report
next to the relevant view.
For more information on reports, see Reports.
Содержание QIAcuity 911000
Страница 2: ...QIAcuity User Manual 06 2022 2...
Страница 68: ...QIAcuity User Manual 06 2022 68...
Страница 126: ...QIAcuity User Manual 06 2022 126 Pressing the Users list button will bring you to the list of all users and their status...
Страница 178: ...QIAcuity User Manual 06 2022 178 Raw image of a well showing dust particles marked with red circles...
Страница 207: ...QIAcuity User Manual 06 2022 207 To open an individual plot click on the appropriate header of the 1D Scatterplot...
Страница 240: ...QIAcuity User Manual 06 2022 240...