QIAcuity User Manual 06/2022
182
Signal map of an image showing areas of bad filling
Crosstalk Correction algorithm
Depending on the optical configuration of your QIAcuity instrument, you can detect up to two or
five fluorescent channels. To compensate the spectral overlap between the fluorescent dyes, a
crosstalk correction algorithm is implemented in the QIAcuity Software Suite. The correction is done
automatically by the software and does not require any user action. The bleed-through signals are
removed from the images and are not taken into account in the result analysis. The crosstalk
correction is correcting an absolute value based on the RFU level of the neighbor channel.
Note
: In case insufficient compensation or overcompensation is seen (e.g., as double negative
bands), check the RFU level of positive signals of neighbor channels if saturated or very bright. By
lowering the RFU level of positive signals, the insufficient compensation or overcompensation might
be reduced.
Содержание QIAcuity 911000
Страница 2: ...QIAcuity User Manual 06 2022 2...
Страница 68: ...QIAcuity User Manual 06 2022 68...
Страница 126: ...QIAcuity User Manual 06 2022 126 Pressing the Users list button will bring you to the list of all users and their status...
Страница 178: ...QIAcuity User Manual 06 2022 178 Raw image of a well showing dust particles marked with red circles...
Страница 207: ...QIAcuity User Manual 06 2022 207 To open an individual plot click on the appropriate header of the 1D Scatterplot...
Страница 240: ...QIAcuity User Manual 06 2022 240...