Note
: For a multi-well selection the user can’t change the thresholds individually. The new threshold
becomes valid for all selected source wells. In case the user would like to change thresholds
individually he has to change the source selection to a single well selection
To reset the software to automatically recalculate the threshold, click
Automatic threshold
.
Again, click
Recalculate
to apply the auto-threshold to the results.
If the user tries to exit the view of the histogram without confirming changes with recalculation
button, a warning modal is shown.
The Histogram tab.
To add any of the histograms to the report, click
Add to report
next to the relevant diagram. For
more information on reports, see section Reports.