QIAcuity User Manual 06/2022
87
Note
: To modify a plate that was already used, you must remove it from the instrument. This ensures
that the plate is unlocked and ready for modifications in the QIAcuity Software suite. If modifications
should be done using the plate configurator, load the Plate again.
Editing the run schedule
Note
: Editing the run schedule is only possible on QIAcuity Eight and QIAcuity Four and for those
users having the appropriate permission (please see Table 9).
When a run starts, it is added to the run schedule and the
Edit schedule
button is shown on the
screen. If the runs are started individually, they will be added to the schedule in the order which
they were started by tapping the
Run
icon on their respective panes. If all the runs are started
at the same time, using the
Run All
button, there is a default order in which the plates are run.
In QIAcuity Eight, the run starts with the first slot in the upper tray and ends with the last slot in the
lower tray. The slot numbers are presented in Table 7.
Table 7. Slot numbers the QIAcuity
Eight
Tray
Slot numbers
Upper
1
2
3
4
Lower
5
6
7
8
In QIAcuity Four, the run starts with slot number 1, and ends with slot number 4. The slot numbers
are presented in Table 8.
Table 8. Slot numbers of the QIAcuity Four
Slot numbers
1
2
3
4
Содержание QIAcuity 911000
Страница 2: ...QIAcuity User Manual 06 2022 2...
Страница 68: ...QIAcuity User Manual 06 2022 68...
Страница 126: ...QIAcuity User Manual 06 2022 126 Pressing the Users list button will bring you to the list of all users and their status...
Страница 178: ...QIAcuity User Manual 06 2022 178 Raw image of a well showing dust particles marked with red circles...
Страница 207: ...QIAcuity User Manual 06 2022 207 To open an individual plot click on the appropriate header of the 1D Scatterplot...
Страница 240: ...QIAcuity User Manual 06 2022 240...