QIAcuity User Manual 06/2022
220
confidence interval. To view detailed information, hold the mouse pointer over its corresponding
bar. A tooltip with detailed information opens.
To view the mean values for replicates on the concentration diagram, click Show mean values for
replicates. If the user clicks into the checkbox to select the mean representation for replicates, the
bars of individual replicates disappear and only one bar is shown at the sample label that represents
the mean mutant fraction value of the replicates. When there are no replicates within selected wells
for mutant target the bars don't change. The corresponding well IDs of the replicates are shown on
the x-axis. This is also applied in same way by default to the view where targets are in different
reaction mixes. In this case individual replicate results are not available. Mean values are calculated
and shown over all replicates that are included in the well selection only. If there are further
replicates of the same sample which are not included in well selection, they are not considered in
the calculated mean value and show their individual results values.
To add the concentration diagram to the report, click
Add to report.
For more information about
reports, see section Reports.
Genome editing
The Genome editing tab contains views which provide insight into the number of edited genomes
in analyzed wells. The analysis results are put into list views, heatmaps, point diagrams, and
concentration diagrams
Note
: Saving genome editing tests is not yet provided Setting up a genome editing analysis
Содержание QIAcuity 911000
Страница 2: ...QIAcuity User Manual 06 2022 2...
Страница 68: ...QIAcuity User Manual 06 2022 68...
Страница 126: ...QIAcuity User Manual 06 2022 126 Pressing the Users list button will bring you to the list of all users and their status...
Страница 178: ...QIAcuity User Manual 06 2022 178 Raw image of a well showing dust particles marked with red circles...
Страница 207: ...QIAcuity User Manual 06 2022 207 To open an individual plot click on the appropriate header of the 1D Scatterplot...
Страница 240: ...QIAcuity User Manual 06 2022 240...