PT610, PT620, PT630
Instruction Manual
59
Example :
Command: 01DD
Response: 01DDA
01DDN
(Command failed)
Comments:
This command only functions if
[331]
= KEY 6.4.3.1 and if W >50e but version 1.02 software still
responds DDN even if successful.
DR
|
Read dynamic weight
Command
:
[ADR][DR]
Response
:
[ADR][DRA][8 digit Weight data]
Example :
Command
: 01DR
Response
: 01DRA 12.364
: 01DRN
(Dynamic weighing could not be read)
: 01DRM--------
(Vesion 1.02 software)
Comments:
The indicator must be in Dynamic mode to read dynamic weight.
Note:
With version 1.02 software the indicator may respond DRM instead of DRA and when
[331]
=
CMIN 6.4.3.1 the response is always DRM-------.
DE
|
Reset to standard weighing
Command
:
[ADR][DE]
Response
:
[ADR][DEA]
DP
|
Write Preset Tare and apply
Command
:
[ADR][DP][+][8 digits PT data]
Response
:
[ADR][DPA]
Example :
Command
: 01DP+000012.4
Response
: 01DPA
01DPN
(Could not be loaded)
Comments:
Preset Tare must be enabled in
[231]
Warning:
the decimal positions of the DP value must match the decimal positions of the instrument
build. i.e. d=0.1, d=0.2, d=0.5 for DP+000012.4
DQ
|
Read Preset Tare
Command
:
[ADR][DQ]
Response
:
[ADR][DQA][8 digit PT data]
Example :
Command
: 01DQ
Response
: 01DQA000123.4
(PT is 123.4)
01DQN
(Could not read PT)
DM
|
Write APW and apply
Command
:
[ADR][DM][12 digits APW data]
Response
:
[ADR][DMA]
Example :
Command
: 01DM000000032648
Response
: 01DMA
01DMN
(Could not write APW)
Comments:
The DM value is sent as 12 digits.
Warning:
For firmware versions 1.02 the APW value will be calculated from the DM value based on the
resolution (d) set during build setup as follows; APW = DM/10
(3-dp)
where dp= number of decimal places
of the division set.
I.E. if the display division is 0000.0 then the APW will be 32648/100 (100 = 10
(3-1)
(i.e. 326.48).
I.E. if the display division is 000.00 then the APW will be 32648/10 (100 = 10
(3-2)
(i.e. 3264.8).
Writing an APW weight will set the scale in counting mode. Use [DE] to reset to basic weighing.