PT610, PT620, PT630
Instruction Manual
58
Command
:
[ADR][DH 1][16 digits name]
␣
Response
:
[ADR][DHA]
Comments:
This sets the active ID1 name, it does not write to ID memory.
Note the position of spaces. Send 16d, fill blank characters with spaces.
DI 1
|
ID1 data write
Command
:
[ADR][DI 1][32 digits data]
␣
Response
:
[ADR][DIA]
Comments:
This sets the active ID1 data, it does not write to ID memory.
Note the position of spaces. Send 32d, fill blank characters with spaces.
DH 2
|
ID2 name write
Command
:
[ADR][DH 2][16 digits name]
␣
Response
:
[ADR][DHA]
Comments:
This sets the active ID2 name, it does not write to ID memory.
Note the position of spaces. Send 16d, fill blank characters with spaces.
DI 2
|
ID2 data write
Command
:
[ADR][DI 2][32 digits data]
␣
Response
:
[ADR][DIA]
Example :
Comments:
This sets the active ID2 data, it does not write to ID memory.
Note the position of spaces. Send 32d, fill blank characters with spaces.
DK 1
|
ID1 name and data read
Command
:
[ADR][DK 1]
␣
Response
:
[ADR][DK 1] [16d name] [32 digits data]
␣ ␣
␣
Example :
Command
: 01DK 1
␣
Response
: 01DK 1 Bolts
M12x1.75P x 40mm
␣ ␣
␣␣␣␣␣␣␣␣␣␣␣␣
␣ ␣
␣␣␣␣␣␣␣␣␣␣␣␣␣␣␣␣
DK 2
|
ID2 name and data read
Command
:
[ADR][DK 2]
␣
Response
:
[ADR[[DK 2] [16d name] [32 digits data]
␣ ␣
␣
Example :
Command
: 01DK 2
␣
Response
: 01DK 2 Nuts
M12x1.75P
␣ ␣
␣␣␣␣␣␣␣␣␣␣␣␣␣
␣␣␣␣␣␣␣␣␣␣␣␣␣␣␣␣␣␣␣␣␣␣␣
DB
|
Write EZAZ data and apply
Command
:[ADR][DB]+[8digits Limit-1]+[8d Lim-2]+[8d Lim-3]+[8d Lim-4]+[8d Lim-5]+[8d Lim-6]
Response
:[ADR][DBA]
Example :
Command
: 01DB+00018.30+00015.30+00020.58+00005.00+00000.00+00000.00
Response
: 01DBA
(Write successful)
01DBN
(Write failed)
Comments:
The indicator must have an option card with relay outputs for the data to save to the indicator.
6 limit values must be sent in all cases, even if only 4 values are required (i.e. check weighing).
Warning:
The decimal positions of the DB value must match the decimal positions of the instrument
build. i.e. d=0.1, d=0.2, d=0.5 for DP+000012.4, even for zero values.
DD
|
Start dynamic weighing
Command
:
[ADR][DD]
Response
:
[ADR][DDA]