51
2-4-8 Counter reset
OFF
ON
OFF
ON
OFF
ON
OFF
ON
OFF
ON
OFF
ON
OFF
ON
T1
T2
T3
T4
T5
*1
*2
SET IN
SET OK OUT
READY OUT
MARKING OUT
D0 to D7 IN
(Counter No.)
TRIG. IN
SELECT 0 to 2 IN
Item
Time
Remarks
T1
10ms or more
Keep the ON status for 10ms or more.
T2
0ms or more
Keep the input until the SET OK is turned to ON.
T3
2 to 510ms
One-shot output. Set the output time on the environment setting screen. There is a
margin for the setting value.
T4
―
•
When laser pumping is completed
Period for creating marking data. It varies depending on the quantity of data to be
marked.
•
When laser pumping is not completed
Period for creating marking data or period until the laser pumping is completed. The
longer one of the two above is employed.
T5
0.5ms or more
After a lapse of 0.5ms or more from specifying SELECT 0 IN to SELECT 2 IN and D0 IN
to D7 IN, turn on SET IN.
*1 : To reset the counter, set SELECT 0 to 2 as follows.
Number input target
SELECT 0 (No. 20)
SELECT 1 (No. 21)
SELECT 2 (No. 22)
Counter reset No.
OFF
OFF
ON
*2 : Specify the counter No. to be reset at D0 to D7. Refer to “Operation of input signal on I/O connector” (P.34) for the detail.
ME-LP400V-EX-6
Содержание LP-400 Series
Страница 12: ...1 Before External Control ME LP400V EX 6...
Страница 24: ...2 Control by I O ME LP400V EX 6...
Страница 56: ...3 Control by Serial Communication RS 232 Ethernet ME LP400V EX 6...
Страница 141: ...Troubleshooting ME LP400V EX 6...
Страница 158: ...Character Code Table ME LP400V EX 6...
Страница 166: ...166 Shift JIS JIS 0 1 2 3 4 5 6 7 8 9 A B C D E F 983F 4F20 984F 4F30 985F 4F40 986F 4F50 ME LP400V EX 6...
Страница 176: ...Index ME LP400V EX 6...
Страница 179: ...No 9000 0062 17V...
Страница 180: ...Panasonic Industrial Devices SUNX Co Ltd 2003 2018 July 2018 9000 0062 17V...