31
Terminal
No.
Name/Description
B4
READY OUT: Trigger ready output
When TRIG. IN (A4) becomes acceptable (the laser radiation becomes ready), this output turns ON.
To turn on READY OUT, the following conditions are required:
•
No error has occurred
•
Laser pumping has completed
•
Internal shutter is open
• The file number switching process has completed
• With the file using “rank function” or “external offset function”, the data number and SET of I/O signals
have been input.
• When the file with the setting to use SIN command is selected, SIN command is transmitted by every
trigger input.
•
When the serial communication control uses the command reception permission (MKM command), the
reception permission state is turned OFF.
B5
MARKING: Marking output
Output ON during marking (laser radiation).
If the marking time is shorter than the specified one-shot output time, MARKING OUT remains ON until the
one-shot output time ends.
B6
MARKING END: Marking end output
Output ON when the marking operation ended.
Even when the marking operation is interrupted due to the alarm, this MARK END turns ON.
This is one-shot output.
Reference
•
READY OUT (B4) is output even during one-shot output of marking end output.
•
MARK END (B6) is output even when marking is interrupted by the Emergency stop switch or laser stop input during
marking.
•
Set the output time of the one-shot duration on the environment setting screen. The setting range is 2 to 510 ms. The
initial setting value is 40 ms. One-shot output time has a small margin of error for the setting value.
B7
LASER OUT: Laser pumping completion output
LASER OUT turns ON for the duration from the completion of the laser pumping until turning OFF the laser
pumping.
The output turns ON approximately 15 to 20 seconds after the laser pumping started.
B8
WARNING: Warning output
Output OFF at warning occurrence.
Contents of warning are indicated by codes in controller file No./error code indicator.
For contents of warning, refer to “Error Indication” (P.151).
B9
ALARM: Alarm output
Output OFF at alarm occurrence. When an alarm occurs, laser pumping turns OFF.
Contents of alarm are indicated by codes in controller file No./error code indicator.
For contents of alarm, refer to “Error Indication” (P.151).
B10
N.C.: No connection
Do not connect externally.
ME-LP400V-EX-6
Содержание LP-400 Series
Страница 12: ...1 Before External Control ME LP400V EX 6...
Страница 24: ...2 Control by I O ME LP400V EX 6...
Страница 56: ...3 Control by Serial Communication RS 232 Ethernet ME LP400V EX 6...
Страница 141: ...Troubleshooting ME LP400V EX 6...
Страница 158: ...Character Code Table ME LP400V EX 6...
Страница 166: ...166 Shift JIS JIS 0 1 2 3 4 5 6 7 8 9 A B C D E F 983F 4F20 984F 4F30 985F 4F40 986F 4F50 ME LP400V EX 6...
Страница 176: ...Index ME LP400V EX 6...
Страница 179: ...No 9000 0062 17V...
Страница 180: ...Panasonic Industrial Devices SUNX Co Ltd 2003 2018 July 2018 9000 0062 17V...