FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H
81
Kinetic Reduction Types
The
Reduction Type
field is located in the
Kinetic Reduction
Configuration
dialog. Available options depend on the software
version in use. A separate software license for ScreenWorks
®
Peak
Pro
TM
adds advanced peak detection and characterization
measurements to the field of standard measurement options (see
). Contact your Molecular Devices Sales Representative for
details.
Configure
Auto-Export
Only available for protocol files, this button opens the
Auto-Export
dialog allowing configuration of
statistics and time sequence files to be automatically
created whenever the protocol is run.
For data files the button changes to Export, and
opens the
File Export
dialog, also opened from the
File > Export
command. See
Configure
Correction
Opens the
Corrections
dialog box to apply various
corrections to modify data display. There is also the
option to view ratiometric data. Settings made here
in protocol files will affect how data are viewed when
the protocol is run, but since all raw data are stored
in data files, these options all remain available for
acquired data as well. See
.
Hide Kinetic
Reduction
Value
Hides the kinetic reduction values that are displayed
in the Multi-Well Graph.
Show Kinetic
Reduction
Value
Shows the kinetic reduction value in each well of the
Multi-Well Graph.
Configure
Kinetic
Reduction
Opens the
Kinetic Reduction Configuration
dialog
to define the parameters used to define the kinetic
reduction.
Reduction Type
—Defines the reduction to be
applied to the kinetic data traces displayed in the
.
Start Read
—Define the first read to be used to
determine the kinetic reduction.
End Read
—Define the last read used to determine
the kinetic reduction.
Read Mode
—Select the read mode to apply the
kinetic reduction to.
Button
Name
Description
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...