FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H
195
Loading Coelenterazine
Conditions to check:
•
Loading buffer (inclusion of FBS or BSA, type of buffer).
•
Dye loading temperature and duration below 22 °C.
•
Coelenterazine specific for Photoproteins.
•
Presence of additives (for example
,
probenecid and pluronic acid
in intracellular calcium assays).
Source Plate
Conditions to check:
•
Type of buffer used should match coelenterazine buffer.
•
For suspension assay, prepare 2x compound directly in read
plate.
•
For adherent assays, prepare a polypropylene plate with 2x
compound.
•
Volume of fluid in source plate in relation to required transfer
volume and plate dead volume.