Troubleshooting
206
0112-0109 H
208: “Unable to load tips.
Tips are already on
pipettor.”
Tips are already
present on pipettor.
If you would like to change
tips, select
Unload Tips
from the
Instrument/Manual
Operation
drop-down
menu. Once tips are
exchanged, select
Load
Tips
from the same
drop-down menu.
If want to keep the present
tips on, clear error and
proceed with your
experiment.
209: “Unable to unload
tips. There are no tips
loaded on the pipettor.”
No tips are present
on the pipettor
head.
Clear error.
233: Unable to dispense.
No tips loaded on pipettor.”
No tips are present
on the pipettor
head.
Load tips manually on the
pipettor head using the
Load Tips
command which
is found un the
Instrument/Manual
Operation
drop-down
menu.
Change protocol to include
loading of tips at the
beginning of the
experiment.
234: “Unable to aspirate.
No tips loaded on pipettor.”
No tips are present
on the pipettor
head.
Load tips manually on the
pipettor head using the
Load Tips
command which
is found un the
Instrument/Manual
Operation
drop-down
menu.
Change protocol to include
loading of tips at the
beginning of the
experiment.
Table 9-4
Troubleshooting the Pipettor (cont’d)
Symptom or Error
Message
Possible Causes
Solutions
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...