Troubleshooting
212
0112-0109 H
218: “Unable to define
plate mask. Please install
515–575 nm emission
filter.”
The 515–575 nm
emission filter is not
installed. Thus, mask
alignment cannot be
performed.
Install the 515–575 nm
emission filter to match
the calibration file.
Calibrate instrument with
new emission filter.
219: “Unable to define
plate mask. Please install
470–495 nm LED banks.”
The 470–495 LED
banks are not installed.
Thus, mask alignment
cannot be performed.
Install the 470–495 nm
LED banks to match the
calibration file.
Calibrate instrument with
new LED bank.
307: “Unable to complete
plate mask definition.
Incorrect number of wells
found. Please verify that
the correct plate is loaded
and repeat.”
Mask alignment failed
because it didn’t find
the correct number of
wells (
e.g.,
must find
384 wells for a 384 well
plate).
Correct read plate format
must be in Read position.
308: “Unable to complete
plate mask definition.
Found overlapping wells.
Please contact Technical
Support. You will not be
able to run protocols
using this plate until the
instrument is repaired.”
Plate is tilted/popped
out of position.
Make sure read plate is
properly aligned in Read
position.
310: “Unable to perform
gain curve. Please insert
515–575 nm emission
filter.”
515–575 nm emission
filter is not installed on
system.
Install the 515–575 nm
emission filter prior to
performing gain curve.
Contact Technical Support
at
+1-800-635-5577.
311: “Unable to perform
gain curve. Please insert
470–495 nm LED banks.”
470–495 nm LED
banks are not installed
on system.
Install the 470–495 nm
LED banks prior to
performing gain curve.
Contact Technical Support
at
+1-800-635-5577.
304: “Gain calibration
failed. Please contact
Technical Support.”
Contact Technical Support
at
+1-800-635-5577.
Table 9-5
Troubleshooting the Optics (cont’d)
Symptom or Error
Message
Possible Causes
Solutions
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...