Troubleshooting
208
0112-0109 H
Some tips are left on the
pipettor after unloading.
Pipettor calibration
failure.
Contact Technical Support
at
+1-800-635-5577.
The tip rack is attached to
the tips.
Tip box failure.
Remove tip box from tips.
Choose
Instrument/Manual
Operation\Unload Tips
to unload tips to tip loading
position.
Contact Technical Support
at
+1-800-635-5577.
242: “The selected read
plate has fewer wells than
the pipettor head. Please
select a plate with a
matching format.”
The pipettor head
and read plate
format is
incompatible (
e.g.,
1536 pipettor head
with 96-well plate).
Load the appropriate
pipettor head type.
Change protocol to include
read plate format that is
compatible with pipettor
head.
243: “The selected source
1 plate has fewer wells
than the pipettor head.
Please select a plate with a
matching format.”
The pipettor head
and Source 1 plate
format is
incompatible (
e.g.,
1536 pipettor head
with 96-well plate).
Load the appropriate
pipettor head type.
Change protocol to include
Source 1 plate format that
is compatible with pipettor
head.
244: “The selected source
2 plate has fewer wells
than the pipettor head.
Please select a plate with a
matching format.”
The pipettor head
and Source 2 plate
format is
incompatible (
e.g.,
1536 pipettor head
with 96-well plate).
Load the appropriate
pipettor head type.
Change protocol to include
Source 2 plate format that
is compatible with pipettor
head.
245: “The selected source
3 plate has fewer wells
than the pipettor head.
Please select a plate with a
matching format.”
The pipettor head
and Source 3 plate
format is
incompatible (
e.g.,
1536 pipettor head
with 96-well plate).
Load the appropriate
pipettor head type.
Change protocol to include
Source 3 plate format that
is compatible with pipettor
head.
Table 9-4
Troubleshooting the Pipettor (cont’d)
Symptom or Error
Message
Possible Causes
Solutions
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...