Running an Experiment
192
0112-0109 H
Experimental Setup Parameter
Fluid Transfer Settings—Aspirate
Parameter
Setting
Read Mode
Luminescence
Excitation Wavelength
None
Emission Wavelength
None
Camera Gain
1
1.
May be adjusted to compensate for cell brightness or saturation problems.
70,000–280,000
Exposure Time
0.53 s
LED Excitation Intensity
NA
Gate Open
100%
Read Interval
1 s
Reads before pipetting
5
Reads during first interval
50–90
Save Assay Images
Possible, but creates very large files
Parameter
Suspension Setting
(pipette cells)
Adherent Settings
(pipette compound)
Cell Flask Spinning Rate Manual setting, Speed 6 NA
Fluid Transfer Type
Single Aspirate, Single
Dispense
Single Aspirate, Single
Dispense
Source Plate
Fill Reservoir speed
Drain Reservoir speed
Drain destination
defaultcellres384
5
4
Cell flask
Source Plate 2
NA
NA
NA
Aspiration Volume
25
μ
L
25
μ
L
Aspiration Height
70
μ
L
5
μ
L
Aspiration Speed
20
μ
L/s
20
μ
L/s
Tip Up Speed
10 mm/s
10 mm/s
0
μ
L 0
μ
L
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...