ScreenWorks Software Overview
58
0112-0109 H
The table below describes the settings in the
Yellow Plate Signal Test
and
Protocol Signal Test
dialog:
Item
Description
Select Plate
Choose the plate type from the drop-down list.
Excitation/Emission
Wavelength
Select the appropriate excitation/emission wavelength
pair for the signal test from the drop-down list.
Note:
Only calibrated excitation/emission
wavelength pairs are displayed in the drop-down
list.
Reading Mode
Select
Fluorescence
or
Luminescence
from the list.
Camera Gain
Select the camera gain for the signal test from the list.
Typical Camera Gain is 1 for the EMCCD camera and 2000
for the ICCD camera, for the 470–495/515–575 nm
excitation/emission pair when reading a yellow test plate.
Gate Open
This option is available only for the Aequorin ICCD
camera. Typical Gate Open values for Fluorescence are
around 6% but can be set to 100%. Luminescence gate
is always 100%, and not user-adjustable.
Excitation Intensity Select the LED intensity for the signal test from the drop-
down list. Excitation intensity is scaled as a percentage of
the total LED output (0–100%). Typical Excitation
Intensity is 80 for the 470–495/515–575 nm
excitation/emission pair when reading a yellow test plate.
Exposure Time
Enter the amount of time (in seconds) to keep the
camera shutter open during the Signal Test. Typical
Exposure Time is 0.4 seconds for the EMCCD camera and
0.53 seconds for the ICCD camera, for the
470–495/515–575 nm excitation/emission pair when
reading a yellow test plate.
Highlight Range
Highlights well values that lie within the set statistical
range. Set the statistical range by using the slider.
Wells Above Range
Displays the number of wells above the statistical range
as determined by the Highlight Range.
Wells Below Range
Displays the number of wells below the statistical range
determined by the Highlight Range.
Maximum
Displays the largest value on the signal test plate.
Average
Displays the average value on the signal test plate.
Minimum
Displays the smallest value on the signal test plate.
Std. Dev
Displays the standard deviation of the signal test plate.
Test Signal
When clicked, initiates a new signal test.
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...