Running an Experiment
172
0112-0109 H
probenecid can be useful in slowing dye leakage from cells, it is toxic to
cells, and therefore inclusion and duration of dye loading step should be
kept to a minimum.
Calcium Assay Kit Experimental Protocol
Cell Handling
The FLIPR
®
Calcium Assay Kit is designed to work with many cell types,
both adherent and non-adherent. We recognize that a variety of cell
handling conditions might be adopted at the discretion of the user,
based on standard operating procedures in the laboratory. In this
section we provide guidelines on how to prepare the cells for use with
the assay kit.
Adherent cells are the most frequently used cells with the kits. They are
typically plated the day prior to an experiment and then incubated in a
5% CO
2
, 37 °C incubator overnight. See the table below for suggested
plating volumes and seeding densities to create an 80–90% confluent
cell monolayer before placing the plates in the FLIPR
®
Tetra
Instrument.
For non-adherent cells, we recommend centrifuging cells from culture
medium and re-suspending the pellet in culture medium on the day of
the experiment. It is recommended after the cells are plated to
centrifuge the plates at 100g for up to 4 minutes (with brake off).
Alternatively, non-adherent cells can be treated like adherent cells,
plating the day before the assay using the same plating volumes and
seeding densities, as long as the cells are seeded onto coated plates
(for example
,
poly-D-lysine or collagen) to ensure good attachment to
the plate bottom.
Preparation of Loading Buffer
The following procedure is designed for preparation of the Loading
Buffer per vial of the Explorer Kit, the Bulk Kit, or the Express Kit.
1.
To prepare the 1X HBSS Buffer plus 20 mM HEPES for the Bulk
and Express Kits only, (Explorer Kit contains ready to use HBSS
buffer plus 20 mM HEPES pH 7.4 in Component B), pipette 100
mL of 10X Hank’s Balanced Salt Solution and 20 mL of 1M Hepes
buffer pH 7.4 into 880 mL cell culture treated water.
Cell Type
(cells/well)
96-well Plate
(100
μ
L
growth
medium)
384-well Plate
(25
μ
L growth
medium)
1536-well
Plate
(4
μ
L growth
medium)
Adherent cells
20,000–80,000
5,000–20,000
1,500–5,000
Non-adherent cells 40,000–200,000
10,000–50,000
3,000–10,000
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...