FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H
269
E112
Door(s) are opened. Close the doors before start your
experiment.
E113
Load Tips position not checked! Check Load Tips position for auto
load tips or manually load tips before running experiment.
E114
The chamber temperature is not in range.
E115
Experiment wash tips and there is no tip washer.
E116
Bottle A is empty! Refill it or use another bottle.
E117
Bottle B is empty! Refill it or use another bottle.
E118
Waste bottle A is full. Please empty the bottle to continue.
E119
Waste bottle B is full. Please empty the bottle to continue.
E120
Remove plates redundant.
E121
Error relate to the data folder. The path is not valid, or cannot be
accessed. Or there are not enough disk space.
E122
Error relate to the export folder. The path is not valid, or cannot
be accessed. Or there are not enough disk space.
E123
The read interval is too short. Please check settings in Read
views.
E124
Invalid aspirate parameters [in the protocol].
E125
Invalid dispense parameters [in the protocol].
E126
The wash tips parameters are invalid.
E127
Invalid fluid bottle specified in wash tips process.
E128
Invalid Read Mode configuration. You tried to take fluorescence
(with excitation) and luminescence (without excitation) reading in
one protocol. This is not allowed.
E129
Invalid Read Mode configuration. Camera gain or gate time is out
of range. Please check your protocol settings.
E130
The protocol has transfer cells and/or wash Cell Reservoir
operations even though the Cell Reservoir is not installed.
E131
Cell Reservoir is not ready.
E132
Cell flask spinning rate could not be set. Please check your
instrument status.
Table A-4
runexperiment command errors (cont’d)
Error
Code
Description
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...