ScreenWorks Software Overview
114
0112-0109 H
Edit Dispense
(continued)
Speed
—Enter speed at which to dispense (in
μ
L/s). Ideal
dispense speed takes into account the volume added, how
fast the signal increases in response to compounds and
strength of the attached cells at the bottom of the plate.
Expel Volume
—Enter an additional volume (in
μ
L) the
pipettor dispenses over the dispense volume. Typically, this
value is equal to the aspirate
Hold Volume
value.
In a multiple dispense, only the last dispense can have an
Expel Volume
.
If
Mix Fluid After Dispense
is selected,
Expel Volume
automatically changes to zero for the dispense associated
with the fluid transfer, to ensure that tips remain in the well
prior to the mix step. However, the expel volume designated
in the dispense is transferred to the
Mix with TF
step, to
help ensure liquid is properly expelled from the tips after
mixing.
Removal Speed
—Enter rate (in mm/s) at which tips are
pulled from well immediately after the dispense. This value is
adjusted for fluid viscosity, fluid volume in the well or if the
transfer is performed to a dry well.
Pause in Well
—Enter time (in seconds) to pause with the
tips in the wells after dispensing and mixing (if specified)
before the next pipettor move.
Read intervals may occur simultaneously with this step.
Typically, the pipettor is paused during a read to prevent
disturbances in the imaging process.
Mix Fluid After
Dispense
Check this box to link a
Mix Fluid
process to the
Transfer
Fluid
process.
This automatically adds a
Mix with TF
step to the protocol,
and opens the configuration page for that step.
See
about mixing options.
Note:
This option is only available when
Fluid
Transfer Type
has a single dispense.
Item
Description
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...