FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H
73
Edit Read Mode
To enable a read mode, click the check box in the far left of the
Set
Read Mode
table. Then double-click in the row or select the row and
press the
Edit Mode
button below the table to configure it. The
Edit
Read Mode
dialog has the following fields:
Item
Description
Read Mode
Name
Enter a name for the selected read mode, if desired.
Reading Mode Select
Fluorescence
or
Luminescence
.
Excitation/
Emission
Wavelength
Select the desired excitation LED and emission filter
wavelength pair from the drop-down list.
Note:
Only calibrated wavelength pairs are available
when the software is in Online mode; see
NONE
instructs the instrument to not turn on any LEDs or
use an empty emission filter position. These are typically used
when performing luminescence experiments.
Excitation
Intensity
Select a value from the list to regulate the intensity of light
emitted by the LED bank for a given Fluorescence read. The
range is 20–100%.
Exposure
Time
Enter a time (in seconds) to regulate the time that light is
collected and measured by the camera. A longer exposure
increases signal intensity.
Exposure time affects the read interval for data collection. For
instance, an exposure time of 0.4 s, added to a required 0.1 s
camera integration time, makes the highest update frequency
0.5 s.
The range is
0.05 s to 30 s
.
Camera Gain
Select a value from the list to regulate the amplification of the
camera power. Increasing the camera gain increases the
signal. The range for the Standard EMCCD camera is 1–240.
With the Aequorin ICCD camera installed, the Fluorescence
read mode is preset to 2000, whereas in Luminescence mode
the range is 2000–280,000.
Camera Gate
This option is only available with the Aequorin ICCD camera in
Fluorescence mode. It further regulates the signal intensity by
selecting the percentage of each frame captured.
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...