Robotic Integration
258
0112-0109 H
Command Parameters
There is one required parameter, temp, for this command.
If a value of zero is used for the temp parameter, temperature control
for the chamber will be turned off. If a valid value for temp is used,
temperature control will be turned on and this value will become the
current setpoint for the chamber temperature. Valid values for temp
are integers between the ambient room temperature plus 5 °C and 40
°C.
Example Command Line
tempcontrolonoff<TAB>12<CR>
Example Response
ok<CR>
Suggested Steps in Command Usage
•
Make sure an experiment is not running.
•
Send
tempcontrolonoff
command.
•
Response will be OK, confirming that command execution has
begun, or a Condition code related to the command indicating
that execution could not begin.
•
Monitor status until the instrument reports that it has reached
the requested temperature.
Command Specific Error Responses
•
Temperature Control System is not functioningC60
Washtips<TAB>Fluid Type, Wash Cycles,
Volume/Stroke, Aspirate Speed, Pump Speed,
Strokes, Hold Time, Dispense Speed<CR>
Command Description
This command is the equivalent of selecting
Instrument
/
Manual
Operation
/
Wash Tips
from the main menu of the ScreenWorks
Software. This selection will wash the pipette tips with the parameters
provided.
Command Parameters
There are eight required parameters for this command. All eight
parameters must be supplied for this command to properly execute. For
some of the parameters, acceptable values may vary depending on the
head type currently installed on the instrument. The following table lists
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...