Running an Experiment
164
0112-0109 H
Setting Up an Assay Protocol
An effective way to reduce time and ensure consistent screening
parameters is to preconfigure protocol (*
.fmp
) files. Some users define
and save protocol files for each cell line, project or user.
Saved protocol files are opened from
File > Open > Protocol File
. In
addition, the last 10 files (6 data and 4 protocol) opened in
ScreenWorks Software are listed in the bottom portion of the
File
menu.
Creating a Protocol File
A protocol file defines experiment parameters:
•
Instrument conditions, for example
,
read mode, plate positions,
file names and file storage.
•
Assay steps, for example
,
fluid transfer and reads.
•
Analysis options, for example
,
groups, corrections and export
parameters.
Protocols are created by an easy drag-and-drop operation that moves
experimental processes from the Process Explorer to the Experiment
window.
Defining the Settings Process
In the
Settings
process of the Experiment window, select the
appropriate parameters for
Read Mode
,
Plate Position
,
Data File
Name
and
Folders
. For a detailed explanation of the
Settings
process
and its options,
.
Creating the Assay Sequence
Select the desired processes from the
Process Explorer
and drag
them to the desired sequence position in the Experiment window.
Click on the process icon you want to define (for example
,
Transfer
Fluid
), and then enter the desired parameters. For more information
on the available protocol processes and their options, see
ScreenWorks Software Overview on page 47
Note:
The protocol processes are executed in a left-to-right order
until there is a combined process of
Transfer Fluid
and
Read with
TF
. After baseline imaging, all pipetting actions and imaging occurs in
parallel.
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...