Running an Experiment
182
0112-0109 H
Recommended Setup Parameters
Recommended experimental setup parameters are provided below.
Note that the addition speeds are faster than in the conventional
protocol because of the increased robustness of the cells after the new
loading procedure.
Faster addition speeds can lead to better mixing of compounds and
lower signal variance across the plate.
Experimental Setup Parameters
Detection Parameters
EMCCD
Camera
ICCD Camera
Read Mode
Fluorescence
Fluorescence
Excitation Wavelength
510–545 nm
510–545 nm
Emission Wavelength
565–625 nm
565–625 nm
Excitation Intensity
50
%
50
Camera Gain
50
2000
Gate Open
—
%
Exposure Length
0.4
1
s
1.
Can be adjusted to increase or decrease the signal if low RFUs or
saturation problems occur.
0.53
Read Interval
1 s
1 s
Addition
Parameters
96-Well
Plate
384-Well
Plate
1536-Well
Plate
Compound
Addition Volume
(
μ
L)
50
12.5
1
Compound
Concentration
(Fold)
5X
5X
X
Dispense Speed
(
μ
L/s) Adherent
Cells
50–100
10–50
5–10
Dispense Speed
(
μ
L/s) Non-
adherent Cells
10–50
5-20
1-6
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...