Fluid Transfer Settings—Dispense
Recommended Assay Settings
Post Assay Cell Suspension Cleaning
1.
The suspension cell system should NEVER be left without
cleaning at the end of an assay day or between assays with
different cells.
2.
Remove cell spinner flask and insert tubing from flask into DI
water solution.
3.
Insert return line tubing to a waste flask making sure that the
tubing does not reach into the waste fluid. It will need to be left
as clean as the inlet tubing.
4.
Perform a manual reservoir wash with water and follow by
wiping out the reservoir with a Kimwipe.
5.
Build a FLIPR
®
Tetra System protocol to wash the reservoir and
all tubing in the following order:
5 times with Endotoxin-free DI water.
5 times with Cytoclense Sterilant wash and soak (2 min).
5 times with Endotoxin-free DI water.
Parameter
Suspension Setting
Adherent Settings
Target Plate
Read Plate
Read Plate
Volume
25
μ
L
25
μ
L
Height
15
μ
L
30
μ
L
Speed
30
μ
L/s
20
μ
L/s
Removal Speed
10 mm/s
10 mm/s
Expel Volume
1
1.
A hold or expel volume will introduce air bubbles into the well causing false
spikes in signal.
0
μ
L 0
μ
L
Pause in Well
None
None
Hold pipettor during
dispense
None
None
Parameter
Setting
Reduction Type
Area under the curve, from or shortly after addition
point to the end of the assay.
Correction
Subtract bias based on Sample 1.
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...