4.5.3
Connecting a load cell with a 6-wire cable
Terminal
Connection/color code
Description
M+
+ Meas./gn
+ Measuring voltage (load cell output)
M-
- Meas./gy
- Measuring voltage (load cell output)
S+
+ Sense/wh
+ Sense voltage
S-
- Sense/bk
- Sense voltage
V+
+ Supply/rd
+ Supply voltage
V-
- Supply/bu
- Supply voltage
Housing
Ground (GND)
Screen (ground)
4 Device installation
X3 Process Indicator PR 5410
EN-47