4.6.14
CC-Link interface
The CC-Link interface card has the type designation PR 1721/35.
The fieldbus card contains all functions to provide a complete CC-Link slave with transfer
rates up to 10 Mbps.
The CC-Link connection is established by the 5-pin terminal
②
.
The card is inserted into "Slot 4"
①
(see Chapter
Note:
This fieldbus card does
not
support the "EasyFill" application.
Technical data
Designation
Data
Transfer rate
156; 625 kbps; 2.5; 5, 10 Mbps
Protocol
CC-Link slave
-
CRC error recognition according to IEC 62026 (EN 50325)
-
Max. 64 station nodes
-
128 I/O bits and 16 (32 bit) words
-
Watchdog Timer
④
(see Chapter
)
Configuration
CSP file "PR1721_1.csp"
Potential isolation
Yes, optocoupler and DC/DC converter
X3 Process Indicator PR 5410
4 Device installation
Minebea Intec
EN-108