7.14.3.2
Modifying a calibration
Note:
[Modify] may be used only for minor changes (e.g. changing the dead load, adapting
mV/V values for dead load and/or Max). Otherwise, always use [New]!
Example:
Resetting the Dead Load
3.
By pressing [Continue] the data are reset to default first (default) before performing
calibration. Press [Cancel] to cancel selection.
4. Determining the maximum load [Max], see Chapter
5.
Determining the scale interval [Scale interval], see Chapter
.
6. Determining the dead load [Deadload at], see Chapter
7.
Calibrating with load [Max at], see Chapter
.
8. Calibrating with mV/V [Max at], see Chapter
9. Calibrating with load cell data (smart calibration) [Max at], see Chapter
10. Carrying out linearization, see Chapter
.
1.
Select
-[Weighing point]- [Calib] and confirm.
2.
Press the [Modify] softkey.
7 Commissioning
X3 Process Indicator PR 5410
EN-161
Minebea Intec