Defining the max (maximum load) and weight unit (Cd 001)
Defining the scale interval (Cd002)
Defining the dead load (Cd003)
nEW
(for new calibration) appears on the display.
7.
Press OK.
The "Cd" calibration menu flashes.
Cd 001
appears on the display.
8. Press OK.
MAX.FSd
(full scale deflection) appears on the display.
9. Press OK.
------.kg
appears on the display.
10. Press
to move the decimal point to the left by one decimal place.
-----.-kg
appears on the display.
11. Press OK.
00300.0 kg
appears on the display (the "3" flashes).
12. Press
(3×) to select the "0" digit.
13. Press
to select another digit to the left.
14. Press
(2×) to select the "2" digit.
15. Press
(5×) to select the weight unit and change it to "g" with
.
16. Press OK.
Cd 002
appears on the display.
17. Press OK.
StEP
appears on the display.
18. Press OK.
1
appears on the display.
19. Press
to change the scale interval to "5" (0.5 g).
20. Press OK.
Cd 003
appears on the display.
21. Press OK.
X3 Process Indicator PR 5410
7 Commissioning
Minebea Intec
EN-128