background image

— Span

— Calibrate with user

weight

— Calibrate with auto

weight

— Adjust with default

weight

Calibrate with standard weight.

— Adjust with internal

weight

Calibrate with internal weight.

— Linearity

Setting linearity.

— Default

Resetting the Device to the Factory Settings:
Accept, ResError, Abort

— User

User defined settings: Accept, ResError, Abort

— Configuration

Configuration of the scale

— Weighing parameters

Weighing parameters

— Ambient conditions

Select environmental conditions: Very stable,
stable, unstable, very unstable

— Application filter

Select application filter: Final readout, Filling
mode, Low filtering, w/o filtering

— Standstill range

Select standstill range: 0.25 digit, 0.5 digit, 1
digit, 2 digit, 4 digit, 8 digit

— Stability symb. delay

Select Stability delay: no delay, short delay,
average delay, long delay

— Tare parameter

Select taring: at any time, not until stable

— Auto zero function

Auto-Zero: auto zero on, auto zero off

— Adjustment function

Adj. sequence: ext.adj.w.fact.wt.,
ext.adj.w.user.wt., ext.adj.w.pres.wt., internal
adjust, ext.lin.w.fact.wt., ext.lin.w.user.wt.,
confirm preload, delete preload, adjust
disabled

— Confirming adjust.

Confirm calibration: manual, automatically

— Zero range

Zero point range: 1% of max load = 1%/Max.
load, 2% of max load = 2%/Max. load, 5% of
max load = 5%/Max. load, 10% of max load =
10%/Max. Load

— Power-On zero

range

Initial zero point range: 2% of max load =
2%/Max. load, 5% of max load = 5%/Max. load,
10% of max load = 10%/Max. load, 20% of max
load = 20%/Max. Load

— Power-On tare/zero

Tare/zero at power on: active, inactive, only for
zeroing

— Measure rate

Measurement speed: normal output, fast
output

— Calibration check

Calibration check: calibration prompt, off

— External adjustment

External adjustment: accessible, blocked

— Application settings

Setting up application

— Application Tare

Application Tare: accessible, blocked

— Number of units

Number of weight units: 1 weight unit, 2 weight
units, 3 weight units

— Weight unit 1…3

Chose weight unit 1…3: Gramm [g], Kilogram
[kg], Carat [ct], Pound [lb], ounce [oz], Troy
ounce [ozt], Tael Hongkong [tlh], Tael

Calibrate with user-defined weight.

Calibrate with automatic weight detection.

7 Commissioning

X3 Process Indicator PR 5410

EN-153

Minebea Intec

Содержание X3 Process Indicator PR 5410

Страница 1: ...10 Instrument Manual Translation of the original instrument manual 9499 050 54103 Edition 11 1 0 10 23 2017 Release 4 50 Minebea Intec GmbH Meiendorfer Str 205 A 22145 Hamburg Germany Phone 49 40 6796...

Страница 2: ...and qualified personnel In correspondence concerning this product the type name and release number serial number as well as all license numbers relating to the product have to be cited Note The produc...

Страница 3: ...r connection 13 2 5 RF interference suppression 13 2 6 Failure and excessive stress 13 2 7 Important note 14 2 8 Repairs and maintenance 14 2 8 1 General information 14 2 8 2 Electrostatically sensiti...

Страница 4: ...ecting load cells of type series PR 6221 48 4 5 6 Testing the measuring circuit 49 4 5 7 External supply to load cells 49 4 5 8 Connecting to input output interface PR 1626 60 50 4 5 9 Connecting an a...

Страница 5: ...ont panel keys 131 7 5 5 Reading out calibration data for WP dead load and max using the front panel keys 133 7 5 6 Search for Pendeo load cells and set dead load using the front panel keys 134 7 5 7...

Страница 6: ...ion 169 7 14 11 Linearization 170 7 14 12 Calculating the test value 170 7 14 13 Saving the calibration 171 7 14 14 Cancelling a calibration 172 7 14 15 Parameter Input 173 7 15 Calibrating xBPI scale...

Страница 7: ...g digital outputs 233 7 17 10 Configuring analog output 238 7 18 System information 239 7 18 1 Displaying the version 240 7 18 2 Displaying the status 241 7 18 3 Showing hardware options 242 7 18 4 Di...

Страница 8: ...ion with DHCP service 283 9 5 3 Updating with a fixed IP address 285 10 ModBus protocol 289 10 1 General description 289 10 2 SPM data in PR 1612 ModBus mode 289 11 SMA protocol 291 11 1 General infor...

Страница 9: ...rror numbers LAST_ERROR 330 16 6 1 Weighing point error 330 16 6 2 Error in the EasyFill application 331 17 Specifications 332 17 1 Note on using free software 332 17 2 Decoding the serial number 332...

Страница 10: ...6 1 Design 335 17 6 2 Weights 335 17 7 Documentation on the CD included 335 18 Appendix 336 18 1 Spare parts 336 18 2 Test connector 337 X3 Process Indicator PR 5410 Table of contents Minebea Intec EN...

Страница 11: ...r preventing hazards are not followed DANGER Warning of personal injury DANGER indicates that death or severe irreversible personal injury will occur if appropriate safety measures are not observed Ta...

Страница 12: ...or the environment may occur if appropriate safety measures are not observed Take appropriate safety measures Note User tips useful information and notes 1 6 Hotline Phone 49 40 67960 444 Fax 49 40 6...

Страница 13: ...The device reflects the state of the art The manufacturer does not accept any liability for damage caused by third party system components or due to incorrect use of the product The use of this produ...

Страница 14: ...supply This device contains electrostatically sensitive components For this reason an equipotential bonding conductor must be connected when working on the open device antistatic protection 2 4 3 Supp...

Страница 15: ...2 Version 24 V DC The device must be connected to the protective grounding conductor The connection can be established via the housing rear panel 2 5 RF interference suppression The device is designe...

Страница 16: ...e and complete description of the problem Maintenance work may only be carried out by a trained technician with expert knowledge of the hazards involved and the required precautions 2 8 2 Electrostati...

Страница 17: ...expanded using the following plug in cards 2 slots Serial input and output board 2 PR 5510 02 or 2 PR 5510 04 1 PR 5510 02 and or 1 PR 5510 04 Analog input and output board PR 5510 07 BCD output boar...

Страница 18: ...32 InterBus S PR 1721 34 DeviceNet PR 1721 35 CC Link PR 1721 36 ProfiNet I O PR 1721 37 EtherNet IP For the internal LAN ModBus TCP Ethernet TCP IP OPC 3 3 Housing 3 3 1 General information The devic...

Страница 19: ...panel cut out The control panel cut out must be made before installing the device all dimensions in mm 3 4 Display and control panel 3 4 1 General information The PR 5410 process indicator can only be...

Страница 20: ...es of up to 7 digits with decimal point and plus or minus sign Available weight units are t kg g or lb lb unit is not permitted as legal for trade in the EU and EEA Above the weight display of the use...

Страница 21: ...lay shows the test value without weight unit No display Test value Gross not tared Positive value Negative value Standstill Zero Description Weight value standstill The gross weight value is within d...

Страница 22: ...range Min Minimum weight t kg g lb These weight units are available 3 4 3 2 Device display The display of the device permits indication of 6 digit weight values digit height 18 mm with decimal point...

Страница 23: ...hing indicates error Symbols mass unit Description Range 1 Range 2 Range 3 t kg g lb These weight units are available 3 4 4 Operating elements 3 4 4 1 User interface The following tables show the basi...

Страница 24: ...entry selection after a confir mation prompt without saving the ch ange Exit parameters menu window Cursor to the right Selection Function keys softkeys Access the Setup menu Depending on the settings...

Страница 25: ...enab led Press the cursor key within an entry to return to the previous character Press the cursor key within an entry to select the next charac ter Within an input pressing the delete key deletes the...

Страница 26: ...d keys are used to navigate through the menus Parameters Use the cursor keys to select the individual parameters Use the key to confirm the selection The required values texts are entered via the alph...

Страница 27: ...the filling process Depending on the settings under Weighingpoint Calib Param Test mode the following is displayed by calling the test with the key later on with Absolute the maximum load with Relativ...

Страница 28: ...he display 3 Press to access the next parameter group CP flashes on the display 4 Press to access the parameter number CP 010 appears on the display The right digit flashes 5 Press to select the digit...

Страница 29: ...13 Press if necessary to select no 14 Press OK to save the change 3 Device description X3 Process Indicator PR 5410 EN 27 Minebea Intec...

Страница 30: ...3 5 Overview of connections X3 Process Indicator PR 5410 3 Device description Minebea Intec EN 28...

Страница 31: ...bits 65 536 parts resolution of 20 000 20 mA For more information see Chapter 4 6 4 Slot 1 or 2 PR 5510 08 BCD output open emitter Outputs 5 digits BCD preceding sign or 3 bytes bi nary Input 1 bit DA...

Страница 32: ...CC Link with up to 10 Mbps For more information see Chapter 4 6 14 Slot 4 PR 1721 36 ProfiNet I O ProfiNet I O with 10 100 Mbit s For more information see Chapter 4 6 15 Slot 4 PR 1721 37 EtherNet IP...

Страница 33: ...cluding cables e g power cable fuse protection load cells junction box data cables console cabinet etc If necessary install the plug in cards device must be disconnected from all voltage sources Follo...

Страница 34: ...terminal on the back of the device 4 3 2 Connecting the equipotential bonding conductor The equipotential bonding conductor must be connected to the grounding terminal on the back of the device 4 4 Ha...

Страница 35: ...nterface card 4 Power supply under the cover 5 Lithium battery under the cover for the power supply 6 Digital inputs and outputs 7 CAL switch 8 Ethernet port 9 RS 232 serial interface 10 Display board...

Страница 36: ...ction RJ 45 socket on the device back Green grn flashing on data traffic activity Yellow yel lights up when there is an existing connection link Transfer rate 10 Mbit s 100 Mbit s full half duplex aut...

Страница 37: ...rate Bit s 300 to 115K2 bit s Data bits 7 8 Input signal level Logic 1 high 3 to 15 V Logic 0 low 3 to 15 V Output signal level Logic 1 high 5 to 15 V Logic 0 low 5 to 15 V Number of signals Input RxD...

Страница 38: ...TxD are switched off 4 4 4 1 Connecting a PR 5110 remote display The remote display PR 5110 can be connected to the device via the internal RS 232 interface or the PR 5510 02 card or the PR 5510 04 ca...

Страница 39: ...10 LInE rS232 oP 12 tokEn oFF oP 13 SEndModE SEnd oP 14 WEIght FolloW oP 15 WPkEy SElEct 4 4 4 2 Connecting a YDP14IS ticket printer The YDP14IS OCEUV ticket printer can be connected via the internal...

Страница 40: ...is Page Mode Press the FEED key to switch from one to the other The procedure can be found in the printer opera ting instructions 4 4 4 3 Connecting an IS platform One IS platform scale with xBPI or S...

Страница 41: ...e electrically isolated by optocouplers and each 2 pin potential free The interface can be configured by software Technical data Description Data Connection Terminal 7 pin Number of inputs 3 CH1 CH2 C...

Страница 42: ...n against incorrect polarity Potential isolation Via optocoupler Cables Screened Connect cable screen wire gauge max 1 5 mm2 to the termi nal terminal contact 14 see example Cable length Max 50 m X3 P...

Страница 43: ...gital outputs The main board is equipped with 3 digital outputs for the process control They are electrically isolated by optocouplers and each 2 pin potential free The interface can be configured by...

Страница 44: ...orrect polarity Max switching current 25 mA Voltage drop 25 mA 3 V Passive external power supply required Potential isolation Via optocoupler Cables Screened Connect cable screen wire gauge max 1 5 mm...

Страница 45: ...ctive load for free wheel diode Supply unit 24 V DC 0 5 V The relay switches when the output is active true To protect the output circuit relays must be equipped with free wheel diodes 4 Device instal...

Страница 46: ...tal outputs The load resistance must be 2 2 1 k Supply unit 24 V DC 0 5 V When the output is active true the output voltage drops from 24 12 V DC to 3 V DC X3 Process Indicator PR 5410 4 Device instal...

Страница 47: ...load cell and connection cables of type PR Color code bk Black bu Blue gn Green gy Gray rd Red wh White For additional information on the connection of load cells and cable junction boxes refer to the...

Страница 48: ...age load cell output M Meas gy Measuring voltage load cell output S Sense Sense voltage S Sense Sense voltage V Supply rd Supply voltage V Supply bu Supply voltage Housing Ground GND Screen ground X3...

Страница 49: ...eas gn Measuring voltage load cell output M Meas gy Measuring voltage load cell output S Sense wh Sense voltage S Sense bk Sense voltage V Supply rd Supply voltage V Supply bu Supply voltage Housing G...

Страница 50: ...supply circuit Load resistance of load cell circuit 75 e g 8 load cells of 650 each The supply voltage is fixed at 12 V DC and protected against short circuits For further technical data see Chapter...

Страница 51: ...5 e g more than 4 load cells with 350 an external load cell supply is required In this case the internal supply is replaced by a potential free external supply The center of the external supply voltag...

Страница 52: ...60 is closed 7 5 V DC the solder bridge 12 on the PR 5410 main board must be opened see Chapter 4 4 1 The sense voltage detection is now switched NOTICE Installation in explosion hazarded areas The s...

Страница 53: ...4 Device installation X3 Process Indicator PR 5410 EN 51 Minebea Intec...

Страница 54: ...20 x 420 The assignments of cable colors are listed in the relevant weighing platform operating instructions Connection designations Combics 1 to 3 devices PR 5410 BR_POS V Supply SENSE_POS S Sense OU...

Страница 55: ...form with 4 wire connection The following links between the terminal contacts are provided from Sense S to Supply V from Sense S to Supply V 4 Device installation X3 Process Indicator PR 5410 EN 53 Mi...

Страница 56: ...onnection 4 6 Accessories 4 6 1 General information The main board has three additional function specific slots These slots can accommodate the following cards X3 Process Indicator PR 5410 4 Device in...

Страница 57: ...13 4 6 14 4 6 15 4 6 16 WARNING Working on the device while it is switched on may have life threatening consequences Before installing or removing a plug in card the device must be disconnected from...

Страница 58: ...possible The connector housings must be conductive containing metal and therefore act as part of the screen They must be screwed firmly into the rear panel as appropriate NOTICE Property damage is pos...

Страница 59: ...an only be used as a point to point connection 2 Loosen and open the cable clip 2 3 Strip the insulation from the cable approx 50 60 mm 4 Cut back the cable screen 3 to 5 mm and bend it over the cable...

Страница 60: ...B Type RS 232 full duplex Transmission rate Bit s 300 19K2 Bit sec Data bits 7 8 Input signal level Logic 1 high 3 15 V Logic 0 low 3 15 V Output signal level Logic 1 high 5 15 V Logic 0 low 5 15 V Nu...

Страница 61: ...his interface card has the type designation PR 5510 04 The plug in card has one RS 232 channel and one RS 422 485 channel which can be used simultaneously and more or less independently depending on t...

Страница 62: ...RS 232 and RS 422 485 Transmission rate Bit s 300 600 1200 2400 4800 9600 19200 bit sec Cable type Twisted pair screened e g LifYCY 3 2 0 20 1 pair of wi res for ground GND Cable gauge 1 5 mm2 preset...

Страница 63: ...ut it has additional signals DCD RI DTR Specifications Description Data Type RS 232 full duplex Data bits 7 8 Input signal level Logic 1 high 3 15 V Logic 0 low 3 15 V Output signal level Logic 1 high...

Страница 64: ...eing installed on the card The RS 485 interface can also be used as a point to point connection Using RS 485 is compulsory with a multi point connection Tristate status Specifications Description Data...

Страница 65: ...gram Factory setting S Function Settings for RS 422 485 1 Tx enable unblock OFF RS 422 ON RS 485 2 Rx enable unblock OFF 4 wire ON 2 wire 3 Rx pull up resistor OFF not connected ON RxB 1K54 V 4 Rx bus...

Страница 66: ...5 1 ON default 2 3 4 5 OFF Last slave RS 485 1 2 3 4 5 ON RS 485 1 3 4 5 ON 2 OFF 4 6 3 2 1 RS 422 point to point connection 4 wire 4 wire transfer mode Full duplex simultaneous sending and receiving...

Страница 67: ...ex simultaneous sending and receiving not possible Switch settings Configuration ON S1 S2 S3 S4 S5 OFF Serial ports parameter Slot 1 2 RS485 Note For more information on the switch settings see Chapte...

Страница 68: ...h settings ON S1 S3 S4 S5 OFF S2 ON S1 S2 S3 OFF S4 S5 Configuration PR 5410 PR 5110 configuration Serial ports parameters Remote dis play Slot 1 2 RS 485 Param Mode single transmitter oP 10 LInE rS48...

Страница 69: ...eral PR 5410 The remote display PR 5110 can be connected to multiple RS 485 interfaces Four wire transmission and full duplex simultaneous sending and receiving possible is possible 4 Device installat...

Страница 70: ...smitters Param Device Id A Param Next Device Id B Serial ports para meter Remote display Slot 1 2 RS485 Param Mode multiple transmitters Param Device Id B Param Next Device Id C Serial ports para mete...

Страница 71: ...interface Four wire transmission and full duplex simultaneous sending and receiving is possible Example Switch settings PR 5410 PR 5110 1 PR 5110 2 PR 5110 3 ON S1 S3 S4 S5 OFF S2 ON OFF S1 S2 S3 S4...

Страница 72: ...SEnd ModE SEnd oP 14 WEIght FolloW oP 15 WPkEy SElEct oP 10 LInE rS485 oP 12 tokEn oFF oP 13 SEnd ModE SEnd oP 14 WEIght FolloW oP 15 WPkEy SElEct 4 6 3 2 6 Connecting an IS platform One IS platform...

Страница 73: ...ells via the xBPI port and the RS 485 interface 2 wire Connections Color code Color Terminal designation Description rd red Supply voltage bu blue Supply voltage gr green B B Signal gy gray A A Signal...

Страница 74: ...o the load cells and junction boxes 4 6 4 Analog inputs and outputs This interface card has the type designation PR 5510 07 The plug in card has one active analog output and four analog inputs The ana...

Страница 75: ...eing installed on the card Analog inputs PIN 2 9 3 10 4 11 5 12 4 channels CH1 GND CH2 GND CH3 GND CH4 GND Analog output PIN 13 14 1 channel 0 4 20 mA Sig GND Note For more information see Chapter 4 6...

Страница 76: ...rror 0 05 Output Max error 0 1 Output Load max 0 500 Output Short circuit protected yes Output Potential isolati on yes Inputs 4 channel current or voltage input 0 20 mA input impedance 250 0 10 V inp...

Страница 77: ...rent 0 20 mA DC Voltage 0 10 V DC Voltage 0 5 V DC S201 1 CH1 ON OFF OFF S201 2 CH2 ON OFF OFF S201 3 CH3 ON OFF OFF S201 4 CH4 ON OFF OFF S201 5 S202 1 CH1 OFF ON OFF S202 2 CH2 OFF ON OFF S202 3 CH3...

Страница 78: ...or Slot 2 The plug in card is used for BCD coded output of a 5 decade weight value Cyclical output of the weight value with 4 bit plus or minus sign stability error The value is intrinsically consiste...

Страница 79: ...7 V Output current Max 50 mA Enable input 5 V 24 V configurable via switch S101 5 V active high 3 1 V active low 1 5 V 24 V active high 16 V active low 10 V Protection against incorrect polarity Poten...

Страница 80: ...5510 08 PIN 2 24 work with a shared power supply at the collector as a reference potential and open emitter outputs A non active output is highly resistant A voltage of approx 1 7 V which is less tha...

Страница 81: ...nductive loads 4 6 5 2 Input The Enable input for PR 5510 08 PIN 25 can control the 23 outputs as DATA_IN As an output data is follow hold tristate and its signal can be configured 5 V TTL 24 V and ac...

Страница 82: ...value with 4 bit plus or minus sign stability error The value is intrinsically consistent and can be output continuously or the digital input DATA_IN control can be used to freeze hold the value Befo...

Страница 83: ...active high 3 1 V active low 1 5 V 24 V active high 16 V active low 10 V Protection against incorrect polarity Potential isolation No Cable type Screened twisted pair e g LifYCY 2x2x0 20 Cable length...

Страница 84: ...6 6 1 Outputs The outputs for PR 5510 09 PIN 2 24 work as a reference potential and open collectors with a shared ground A non active output is highly resistant A voltage of 0 9 V is applied to an ac...

Страница 85: ...nal external free wheel diode additional internal free wheel diode only for very small inductive loads 4 6 6 2 Input The Enable input for PR 5510 09 PIN 25 can control the 23 outputs as DATA_IN As an...

Страница 86: ...nternal free wheel diode ON do not activate internal free wheel diode OFF Switch S102 Input PIN 25 Function for the output data S102 for 1 2 3 4 5 DATA_IN follow hold PIN 2 24 OFF OFF ON ON or OFF ON...

Страница 87: ...a modification on the output memory may occur at the same time that the external request signal changes from data hold to data valid This means that the requested device still has to wait 100 s until...

Страница 88: ...ules are always enabled PIN 25 DATA_IN is an output 4 6 9 Opto decoupled inputs and outputs This interface card has the type designation PR 5510 12 The card converts the external binary signals from t...

Страница 89: ...l power supply required Input Current 7 mA 24 V 3 mA 12 V protected against incorrect polarity Output Qty 12 Output Voltage Max switching voltage 32 V DC Max switching current 25 mA Voltage drop 25 mA...

Страница 90: ...ir e g LifYCY 2x2x0 20 Cable length max 50 m screened Scope of delivery D Sub male connector 37 pin Incl shielded housings Accessories PR 1623 10 connection cable 4 m PR 1623 20 relay unit PR 1623 30...

Страница 91: ...red I O channels potential free no shared reference Outputs Inputs Digital inputs Channel CH1 CH2 CH3 CH6 Assignment Tare reset tare po sitive flank Print command positive flank Not used 4 Device inst...

Страница 92: ...he zero setting range Output limit 1 Output li mit 2 Gross weight value is within d of zero Scale er ror ADC error Not used Input circuitry Contact input Output circuitry Voltage output The load resis...

Страница 93: ...Current output additional free wheel diode for inductive loads 4 Device installation X3 Process Indicator PR 5410 EN 91 Minebea Intec...

Страница 94: ...4 6 9 1 Circuit diagrams PR 5510 12 PR 1623 10 PR 1623 20 X3 Process Indicator PR 5410 4 Device installation Minebea Intec EN 92...

Страница 95: ...green Module is initialized and ready for operation Flashing 2 Hz green Module is not initialized Flashing 1 Hz red RAM check error Flashing 2 Hz red Check error in ASIC and FLASH ROM Module is defect...

Страница 96: ...ot 4 see Chapter 4 6 1 Technical data Description Data Transfer rate 9 6 kbit s to 12 Mbit s auto baud rate detection Connection mode ProfiBus network connections can be made released without affectin...

Страница 97: ...L Note The GSD file is stored on the CD supplied with the device fieldbus directory of the respective device The current file is also available to download online http www minebea intec com ProfiBus c...

Страница 98: ...se plug connections with integrated terminating resistors 4 6 11 1 Controls on fieldbus card The terminating resistors can be switched on ON and off by pressing the bus ter mination switch see Chapter...

Страница 99: ...tions in the bus cable Bus termination switch is located in the module cover and can be accessed from outside Bus termination switch ON The bus termination is switched on If the module is the last or...

Страница 100: ...4 6 11 4 Connection diagram for a master with three slaves X3 Process Indicator PR 5410 4 Device installation Minebea Intec EN 98...

Страница 101: ...inserted into Slot 4 see Chapter 4 6 1 Specifications Description Data Transmission rate 500 kbit s or 2 Mbit s switchable Protocol Interbus S master slave Fixed telegram length Deterministic cyclical...

Страница 102: ...Cable sheath pea green Special InterBus cable certified Housing S Screen 1 DO1 yellow not inverted data output 2 DI1 gray not inverted data input 3 GND brown signal ground 4 not connected 5 GND signa...

Страница 103: ...1 LEDs in the module cover The module cover can be found at the rear of the device LED 1 CC RC LED 2 BA LED 3 RD LED 4 TR Off Constant green Cable OK no reset mode in the master Bus is active PCP comm...

Страница 104: ...the supply voltage 4 6 12 3 Connection diagram for a master with three slaves NOTICE At bus OUT pins 5 and 9 must be bridged if another slave follows See connection diagram X3 Process Indicator PR 541...

Страница 105: ...4 Device installation X3 Process Indicator PR 5410 EN 103 Minebea Intec...

Страница 106: ...escription Data Transfer rate 125 250 and 500 kbit s Protocol DeviceNet Master Slave Polling procedure polled IO CRC error recognition according to IEC 62026 EN 50325 Max 64 station nodes Data width m...

Страница 107: ...the CD supplied with the device fieldbus directory of the respective device The current file is also available to download online http www minebea intec com DeviceNet terminal Allocation of the 5 pin...

Страница 108: ...cover can be found at the rear of the device LED 1 No function LED 2 Network status LED 3 No function LED 4 Module status Off No power Constant green Link detected online and connected Module is worki...

Страница 109: ...4 6 13 3 Connection diagram for a master with three slaves 4 Device installation X3 Process Indicator PR 5410 EN 107 Minebea Intec...

Страница 110: ...4 see Chapter 4 6 1 Note This fieldbus card does not support the EasyFill application Technical data Designation Data Transfer rate 156 625 kbps 2 5 5 10 Mbps Protocol CC Link slave CRC error recogni...

Страница 111: ...th the device Fieldbus directory of the respective device The current file is also available for download via the Internet http www minebea intec com CC Link terminal Allocation of the 5 pole terminal...

Страница 112: ...ction HW No power supply Normal function No power supply No transmission data No power supply No data recepti on Constant green Normal function Send data Receive data Constant red CRC error No admissi...

Страница 113: ...DML Vx xx Sartorius PR5410 xxxxxx xml Potential isolation Yes Cable type Twisted pairs screened e g patch cable CAT5 Autolink stra ight or crossover Cable impedance 150 Cable length to HUB Max 115 m C...

Страница 114: ...ngs and activate the master as a DHCP server W Allocate IP adr via IO controller NOTICE Slave master device names A unique device name must be assigned out of the master This name is given highest pri...

Страница 115: ...error no station name IP address internal er ror 4 6 16 EtherNet IP interface The EtherNet IP interface card has the type designation PR 1721 37 The fieldbus card is equipped with a standard RJ 45 soc...

Страница 116: ...w odva org Industry compatible CE UL and cUL Note The EtherNet IP card is supported by PR 5410 firmware release 1 30 or higher The IP address and subnet mask are set under Fieldbus parameters see also...

Страница 117: ...trolled by scan ner Online link estab lished Flashing green Not configured or scanner not active Packet is received or transmitted Online no link es tablished Constant red Fatal error Duplicate IP add...

Страница 118: ...is not possible 5 1 2 Display functions Display of gross net or tare weight Tare reset tare Set gross to zero Print weight Display of weight values or remote display Functions via digital inputs and...

Страница 119: ...tion 6 1 Functions 6 1 1 General information The EasyFill application is used for single component filling The application allows for quick and reliable filling and emptying of containers The filling...

Страница 120: ...8 alphanumeric characters Set point Set point Input weight adopt unit from the calibration Preset Preset point for switching from coarse flow to fine flow Input weight adopt unit from the calibration...

Страница 121: ...l configuration and interface data are stored in non volatile EAROM memory Unauthorized data changing can be prevented by an access code The front panel keys can be disabled Additional write protectio...

Страница 122: ...y will show Error 15 if the CAL switch is closed The position of the CAL switch is shown with VNC WEB browser under Show status opened opened no write protection closed closed write protection is acti...

Страница 123: ...arting has the following effects on the device Current process steps are deleted The device is reset to its factory settings The network settings are not changed 7 2 Switching on the device The device...

Страница 124: ...7 2 1 Setting the date and time Entering the date and time via Notebook PC VNC WEB browser See Chapter 7 17 2 Entering the date and time using the front panel keys 1 Press SEtuP appears on the display...

Страница 125: ...M legal for trade 015 StStIM Standstill time 7 3 Switching off the device The device is switched off disconnected from power supply by pulling the plug 7 4 Warm up time The device requires a warm up...

Страница 126: ...3oF Action condition do Digital outputs settings via VNC for Standard application only 040 bCd out BCD mode GroSS NEtIt net if tared otherwise gross SELECt trnSP 041 outP 1 Output 1 042 outP 2 Output...

Страница 127: ...ent 085 Et Addr Hardware address MAC ID This address is also located on the label on the side of the device see Chapter 7 9 dt Date and time 086 YEAr YYYY dAtE MM TT tIME HH MM VS Program versions 087...

Страница 128: ...160 nuMPrt Number of printouts 161 SEqnr Enter sequence number 162 PrtMod Select printing mode LAyout appl layout or nICELb NiceLabelExpress 163 168 LAyou 1 6 Configure line 1 6 nEt net weight GroSS...

Страница 129: ...et the dead load with empty scale Note The device has the following default setting when recalibration is started Max 3000 kg Scale interval 1 kg Start Recalibrating Cd 000 1 Switch on the device see...

Страница 130: ...the left by one decimal place kg appears on the display 11 Press OK 00300 0 kg appears on the display the 3 flashes 12 Press 3 to select the 0 digit 13 Press to select another digit to the left 14 Pre...

Страница 131: ...dead load appears on the display 27 Press OK 00000 0 kg appears on the display The dead load is set 28 Press OK Cd 004 appears on the display 29 Press OK SPAn appears on the display 30 Press OK LoAd r...

Страница 132: ...Cd 005 appears on the display 38 Press Exit SAVE appears on the display 39 Press OK YES appears on the display 40 Press OK SAVE appears on the display during the save process 41 Press Exit to exit the...

Страница 133: ...ale 11 Press OK A weight value for the dead load appears on the display 12 Press OK 00000 0 kg appears on the display The dead load is set 13 Press OK Cd 004 appears on the display 14 Press Exit SAVE...

Страница 134: ...ears on the display 7 Press OK MAX FSd full scale deflection appears on the display 8 Press OK The set maximum load appears on the display 9 Press OK 10 Press OK Cd 002 appears on the display 11 Press...

Страница 135: ...ears on the display 18 Press OK Cd 004 appears on the display 19 Press OK SPAn appears on the display 20 Press OK The mV V value for the maximum load appears on the display 21 Press OK Cd 005 appears...

Страница 136: ...e display 3 Press to change the parameter number 000 flashes 4 Press multiple times to select parameter number 003 5 Press OK dEAdLo appears on the display 6 Press OK LoAd appears on the display 7 Pre...

Страница 137: ...h for connected load cells is conducted and the dead load is set immediately thereafter 7 Press Exit to exit the Setup menu 1 Press SEtuP appears on the display 2 Press OK The Cd calibration menu flas...

Страница 138: ...to enter the PIN code 14 Press OK If a wrong PIN code has been entered PInWro appears on the display 15 Press OK 000000 appears on the display 16 Press the cursor keys to enter the correct PIN code Th...

Страница 139: ...The Cd calibration menu flashes Cd 000 appears on the display 7 Press multiple times to select oP oP 060 appears on the display 8 Press to change the parameter number 000 flashes 9 Press to select the...

Страница 140: ...less significant part of the address 000 000 appears on the display 11 Press the cursor keys to select the desired numbers 12 Press OK FP 098 appears on the display 13 Press OK SubnEt subnet mask app...

Страница 141: ...Exit to exit the Setup menu 1 Press SEtuP appears on the display 2 Press OK The Cd calibration menu flashes Cd 000 appears on the display 3 Press multiple times to select nP 080 4 Press OK dHCP DHCP...

Страница 142: ...play 15 Press the cursor keys to select and set the desired numbers 16 Once the last number is selected and set press to set the less significant part of the mask 255 255 appears on the display 17 Pre...

Страница 143: ...r keys to select and set the desired numbers 30 Press OK nP 085 appears on the display 31 Press OK EtAddr hardware address MAC ID appears on the display 32 Press OK The more significant part of the ad...

Страница 144: ...ress OK The Cd calibration menu flashes Cd 000 appears on the display 3 Press multiple times to select nP 080 4 Press OK dHCP DHCP appears on the display 5 Press OK on appears on the display 6 Press O...

Страница 145: ...e desired sequence number 7 Press OK YEAr appears on the display 8 Press OK The year of the alibi entry appears on the display 9 Press OK dAtE appears on the display 10 Press OK The date of the alibi...

Страница 146: ...elationship is checked cyclically at 2 3 minute intervals no appears on the display 17 Press to select YES 18 Press OK The alibi entry is printed out 19 Press Exit to exit the Setup menu 1 Press SEtuP...

Страница 147: ...For this the program must be installed and started on a notebook PC 1 On the notebook PC set the LAN local and Internet Protocol properties to Obtain an IP address automatically depending on the oper...

Страница 148: ...the device Regular running light in LED 1 2 3 4 Click the button to re start the network search run Waiting 2 3 minutes is essential 5 Acoustic signal for each device that was detected as online Note...

Страница 149: ...orkstation PC 7 11 Operation using VNC VNC on the enclosed CD ROM stands for virtual network computing and is a program for remote operation of computers The program distinguishes between the VNC serv...

Страница 150: ...wer the web browser e g Microsoft Internet Explorer Mozilla Firefox Web browser etc can be used directly The disadvantage is that an additional Java installation is required In addition to VNC this in...

Страница 151: ...ll For description of the web menu see Chapter 9 2 1 NOTICE If the VNC viewer is closed on the setup level the device reboots and the web menu is not accessible for several seconds If the web menu and...

Страница 152: ...rval unit Unit of the printing interval Selection Seconds Minutes Hours Meas time Printlayout Item 1 6 Print layout for line 1 6 see Chapter 7 17 4 Fieldbus protocol HW address MAC ID display e g 00 9...

Страница 153: ...00001 3000 999999 kg t lb oz g mg Scale interval Input of the scale interval 1 2 5 10 20 50 Dead load at 0 000000 mV V or by load by load 0 00001 999999 kg t lb oz g mg Max at 1 000000 mV V or by loa...

Страница 154: ...ax without error message 0 999999 d Minimum weight Input of the minimum load 0 d 50 d 999999 d Range mode Range selection Single range Multiple range Multi interval See also Chapters 7 14 15 2 and 7 1...

Страница 155: ...w user wt ext adj w pres wt internal adjust ext lin w fact wt ext lin w user wt confirm preload delete preload adjust disabled Confirming adjust Confirm calibration manual automatically Zero range Zer...

Страница 156: ...s after 50 updates after 100 updates Parameter change parameter change can be changed cannot be changed Select specification group Specif group 1 6 Select specification group of the scale see operatin...

Страница 157: ...g t lb g mg oz Scale interval Enter verification interval 1 d 1 2 5 10 20 50 displayed according to the decimal places at Max and the weight unit Dead load dead load Weight of the empty scale CAL weig...

Страница 158: ...ion interval Only for Multiple range or Multi interval LC name Give each load cell a name LC 1 n load cell 1 n e g PR6224 xx Input of max 20 alphanumeric characters Default Settings are reset to facto...

Страница 159: ...onfigure inputs Action Action selection see Chapter 7 17 8 1 Condition Condition selection see Chapter 7 17 8 2 Analog mode Analog output parameter Gross D08 gross Net if tared D09 Net if tared Select...

Страница 160: ...point 7 14 1 General information For legal for trade application select Weighing point Calib Param where the W M must be set to OIML before calibration is started see Chapter 7 14 15 The calibration...

Страница 161: ...displayed 7 14 2 2 Increased resolution 10 fold In the Weighing point Calib menu the weight is displayed with 10 fold resolution also with the CAL switch closed with After 5 seconds the display return...

Страница 162: ...lues for dead load preload and or Max changing the scale interval Otherwise select the New menu item Under Weighing point Calib choose between New and Modify 7 14 3 1 Performing a new calibration 1 Se...

Страница 163: ...Determining the maximum load Max see Chapter 7 14 4 5 Determining the scale interval Scale interval see Chapter 7 14 5 6 Determining the dead load Deadload at see Chapter 7 14 6 7 Calibrating with loa...

Страница 164: ...y to reset the dead load 5 Press to exit the calibration A prompt window opens 6 Press the Yes softkey to close the menu without calculation of the test value 7 Press to exit the calibration for good...

Страница 165: ...with or without a decimal point Note Error messages calibration see Chapter 16 4 7 14 5 Determining the scale interval The scale interval d is the difference between two successive display values Wit...

Страница 166: ...Dead load at line is selected Cannot be changed here while linearization is active Changes cannot be made while linearization is switched on Only deleting of the linearization points deactivates the...

Страница 167: ...arization was carried out see Chapter 7 14 11 after selection of the line Max at the following tip is displayed Cannot be changed here while linearization is active Changes cannot be made while linear...

Страница 168: ...nto account The STAR load cell data is based on the acceleration of gravity in Hamburg Germany 9 81379 m s2 1 Press the by load softkey 2 Place the CAL weight on the scale 3 Enter the weight of the CA...

Страница 169: ...he dead load rather than the maximum load using the formula specified earlier 1 Calculating SPAN value for Max and if necessary for the dead see Chapter 7 14 8 1 2 Press the by mV V softkey 3 Entering...

Страница 170: ...with nominal characteristic value Cn 2 mV V at maximum capacity of 2000 kg Maximum load 1000 kg Dead load 500 kg Load cell supply voltage 12 V DC 7 14 9 Calibrating with load cell data smart calibrat...

Страница 171: ...stance When all LC same is set only one value for the characteristic value LC sensitivity and the output resistance LC resistance are required For each LC specific press the Enter softkey to enter ind...

Страница 172: ...r 7 14 11 the following note appears on the display after the Dead load at line is selected Cannot be changed here while linearization is active Changes cannot be made while linearization is switched...

Страница 173: ...ive the deviation from the test value 7 14 13 Saving the calibration Quit calibration by pressing the softkey You are prompted to confirm whether calibration should be closed without determining the t...

Страница 174: ...ibration by pressing the softkey You are prompted to confirm whether calibration should be closed without determining the test value If not all data was determined when calibrating with New e g dead l...

Страница 175: ...e The duration of a measurement can be selected Selection 5 ms 10 ms 20 ms 40 ms 80 ms 160 ms 320 ms 640 ms 960 ms 1280 ms 1600 ms Digital filter Selection of the digital filter filter characteristic...

Страница 176: ...s are connected to an external supply it is possible to switch to 8 V to tune Sense voltage monitoring to the lower supply voltage Selection below or equal 8 V 8 V above 8 V 8 V Test mode With absolut...

Страница 177: ...rology a value 2 of the Max must be entered example 60 d for 3000 e of class III Zerotrack range Range within which automatic zero tracking compensates deviations Setting range 0 25 10000 00 d In lega...

Страница 178: ...eight without weight unit is shown 7 14 15 2 Multiple range scale Class III or single range scale Class I and II with variable scale interval The multiple range scale is a scale with two or more weigh...

Страница 179: ...ervals Each interval range has a different scale interval where the weighing range is automatically switched depending on the load on the scale and also when the load is placed on removed from the sca...

Страница 180: ...50 Note During calibration the multi interval function is always switched off Example Range mode Multi interval Interval range 1 0 1500 kg when calibrating set scale interval 1 kg Interval range 2 15...

Страница 181: ...PI scale is not possible 7 15 2 Parameters for serial interface 1 Select Serial ports parameter xBPI Port and confirm The following window opens 2 Select the desired interface and confirm 3 Press the...

Страница 182: ...e in bus operation mode Serial number of the xBPI scale or weighing mode with legal for trade application 4 If necessary change the parameters Only the baudrate and stop bits can be set for an xBPI sc...

Страница 183: ...gy With serial number 0 checking is omitted Setting range 0 99999999 SBN Address When the address is not set to 0 bus operation is active Possible addresses 1 31 i e max 31 xBPI scales can be operated...

Страница 184: ...following window opens The parameters of the xBPI scale are read into the device 3 Select Show device info with the cursor and confirm The following window opens 4 Change the user ID and SBN address...

Страница 185: ...cted scale are displayed Press No for exit from the menu without data change 6 Select Select group of specification using the cursor and confirm 7 Select and confirm the desired specification group 8...

Страница 186: ...meters are listed as an overview in the following see Chapter 7 15 5 3 Note Only the parameters supported by the connected scale are displayed Press No for exit from the menu without data change 7 15...

Страница 187: ...x load Power On tare zero active inactive only for zeroing Measure rate normal output fast output Calibration check Off Calibration prompt External adjustment Accessible Blocked Maximum capacity reduc...

Страница 188: ...ed one level lower two levels lower three levels lower 1 0 5 0 2 0 1 0 05 0 02 0 01 Multi interval increased by 10 Communication type SBI protocol xBPI protocol Baudrate for SBI 150 baud 300 baud 600...

Страница 189: ...te with ID 22 byte Data output interval with each display after 2 updates after 5 updates after 10 updates after 20 updates after 50 updates after 100 updates Parameter change can be changed cannot be...

Страница 190: ...d into the device 3 Select Calibration with the cursor and confirm The following window opens 4 For setting the dead load remove the weight from the scale and select Set using the cursor and confirm A...

Страница 191: ...previously stored user weight is displayed 4 Change the weight value if necessary using the keyboard and confirm The calibration process is carried out without a weight The calibration status is disp...

Страница 192: ...5 2 The xBPI scale weighing point has been selected see Chapter 7 15 3 The platform has been set up see Chapter 7 15 4 1 Select Weighing point xBPI Scale and confirm 2 Press the Setup softkey The para...

Страница 193: ...nd platform is active Procedure 1 Select Weighing point xBPI Scale and confirm 2 Press the Setup softkey The parameters of the xBPI scale are read into the device 3 Select Calibration Adjust with defa...

Страница 194: ...was set to manual The communication between the device and platform is active Procedure 2 Press the Setup softkey The parameters of the xBPI scale are read into the device 3 Select Calibration Adjust...

Страница 195: ...arity Default using the cursor and confirm The first linearization point to be calibrated is displayed 4 Place the displayed weight on the scale The deviation is displayed in the last line with increa...

Страница 196: ...PR 5510 04 interface card must be installed in the device see Chapter 4 6 1 Select Serial ports parameter xBPI Port and confirm The following window opens 2 Select the desired interface and confirm 3...

Страница 197: ...e followed Search for load cells see Chapter 7 16 5 Assign load cells see Chapter 7 16 6 4 Select Baudrate and confirm A selection window opens 5 Select 19200 bd and confirm 6 Select Stopbits and conf...

Страница 198: ...necessary see Chapter 7 16 10 3 Note For further information about calibrating weighing points see Chapter 7 14 3 7 16 5 Searching load cells 1 Select Weighing point Weighing point A 2 Press the Assig...

Страница 199: ...start a new search process Type Type of load cells No of LC Number of load cells LC 1 n Serial number name of the load cells WP serial number Weighing point serial number is displayed after search A w...

Страница 200: ...gn softkey You are prompted to confirm 3 Press the Continue softkey to reset the dead load information and start the assignment procedure 4 Press the Cancel softkey to not start the assignment 5 Place...

Страница 201: ...ct php Number of platforms only for Pendeo Truck load cell This parameter is shown only in the case of 8 load cells Entering the number of platforms Number of vessel feet only for Pendeo Process load...

Страница 202: ...is example 0020 The scale interval d is calculated based on the maximum weight value Dead load To use the empty scale as dead load normal case Do not load scale Press the by load softkey Note If the d...

Страница 203: ...Weighing point Weighing point A Assign Service The service window appears Item number serial number dead load and current weigh of connected load cells are displayed 7 16 9 1 Deactivating the load ce...

Страница 204: ...faulty load cell and confirm to deactivate the cell 2 Press the Accept softkey The warning symbol replaces the weight unit 1 After inserting and connecting the new load cell select the line of the de...

Страница 205: ...oftware corner correction If the corners are loaded in succession the same value should be displayed on the device at all times An excessive deviation almost always means that the scale is tilted or i...

Страница 206: ...W M Setting for legal for trade mode Note The transmitter of the series PR 5410 are not approved for calibration Unbal Check deviat The plausibility check is activated when the average deviation is 0...

Страница 207: ...rology a value 2 of the Max must be entered example 60 d for 3000 e of class III Zerotrack range Range within which automatic zero tracking compensates deviations Setting range 0 25 10000 00 d In lega...

Страница 208: ...s again to return to the previous state Note The scale must not be loaded If the entire zero setting range is already utilized you can still correct the dead load subsequently without affecting other...

Страница 209: ...is only available if under Operating parameter Application Standard has been selected Configuring digital inputs and outputs Digital i o parameter Note This menu item is only available if under Opera...

Страница 210: ...rinter and confirm A selection window opens 2 Select the desired interface and confirm The selected interface is displayed 3 Press the Param softkey to set the parameters The following window opens X3...

Страница 211: ...Press to return to the previous window 7 Press the Config softkey to define the print settings The following window opens 8 Select parameters and confirm 9 Select and confirm the desired printer sett...

Страница 212: ...audrate and confirm A selection window opens 5 Select the desired transmission speed and confirm 6 Select Mode and confirm 7 If several remote displays are connected select the multiple transmitters m...

Страница 213: ...lected interface is displayed 3 Press the Param softkey to set the parameters The following window opens 4 Select Baudrate and confirm A selection window opens 5 Select the desired transmission speed...

Страница 214: ...lave ID and confirm 9 Enter and confirm the slave address in this case 65 10 Press 2 to exit the menu and save 1 Select SMA and confirm A selection window opens 2 Select the desired interface and conf...

Страница 215: ...on speed and confirm 6 Press two times to exit the menu and save 1 Select EW COM and confirm A selection window opens 2 Select the desired interface and confirm The selected interface is displayed 3 P...

Страница 216: ...ection window opens 7 Select the desired transmission speed and confirm 8 Select Bits and confirm A selection window opens 9 Select the desired bit quantity and confirm 10 Select Slave ID and confirm...

Страница 217: ...eters The following window opens 4 Select Baudrate and confirm A selection window opens 5 Select the desired transmission speed and confirm 6 Select Stopbits and confirm A selection window opens 7 Sel...

Страница 218: ...1 Select Date Time and confirm The following window opens 2 Select the individual digits and use the keypad to overwrite them and confirm 3 Select the individual digits and use the keypad to overwrit...

Страница 219: ...12223 Use alibi memory Using the alibi memory The weight values stored in the alibi memory are defined here none do not store any data records Gross Net Gross Net Tare Gross Net Gross Tare Sequence nu...

Страница 220: ...or automatically toggle with reset tare on zeroset to the gross mode If the zero setting key with these settings has no effect the configured zero setting range around the zero point set with the dead...

Страница 221: ...ments will be printed Gross weight Net weight Tare weight Date Time printed in format DD MM YYYY HH MM SS Sequence number counter for individual print orders max 6 digits 000001 comes after 999999 CR...

Страница 222: ...Open the menu via Fieldbus parameter This menu item can only be selected if a fieldbus card is installed The protocol displayed automatically depends on of the installed fieldbus card ProfiBus DP for...

Страница 223: ...S7 Requirements PR 1721 31 ProfiBus DP is installed The parameters are selected and saved Procedure Note Further details can be found in the supplementary application manual How to available upon requ...

Страница 224: ...8 bytes counted from byte 0 7 7 17 5 3 ProfiNet I O settings for S7 Requirements PR 1721 36 ProfiNet I O is installed Procedure Note Further details can be found in the supplementary application manua...

Страница 225: ...pter Description of the I O Area Example The gross weight should be read I O size 8 bytes counted from byte 0 7 7 17 6 Network parameters Under this menu item you can define the network parameters for...

Страница 226: ...If is checked presettings DHCP selected the server automatically allocates the IP address subnet mask and default gateway If is not checked the settings IP address Subnetmask and Default gateway must...

Страница 227: ...the same principle The limit values always refer to the gross weight For the SPM addresses for the limits see Chapter 13 4 NOTICE The limit values of an xBPI weighing point are scale specific The sca...

Страница 228: ...a hysteresis of 10 kg In the event of a power failure both outputs turn to off OFF thus indicating underfill and overfill simultaneously Example 2 If the Limits 1 and 2 are the same for On and Off on...

Страница 229: ...ear marker 1 clr marker 2 X65 0 Clear marker 2 clr marker 3 X66 0 Clear marker 3 Note The limit values can be assigned to the outputs directly in the I O parameters Markers can be set for all limits i...

Страница 230: ...al error in weighing point not ac tive no error above Max X33 0 Weight above Max not active overload X34 0 Weight above Max plus the overload value not active below zero X35 0 Weight not below zero ce...

Страница 231: ...2 active actual limit 3 X18 1 Limiting signal 3 active ADC error X32 1 General error in the weighing point above Max X33 1 Weight above Max overload X34 1 Weight above Max plus the overload value belo...

Страница 232: ...nked with conditions that must be met before an action can be started The parameters for the digital inputs are defined under Digital i o parameter 1 Highlight and confirm the condition line of the ap...

Страница 233: ...Description no action no function set marker 1 X64 1 Set marker 1 set marker 2 X65 1 Set marker 2 set marker 3 X66 1 Set marker 3 select net X72 1 Select net set zero X112 1 Set zero set tare X113 1...

Страница 234: ...17 8 2 Determining a condition The selected action of each digital input can be combined with a condition that must be met for signal change from 0 to 1 on or for signal change from 1 to 0 off The co...

Страница 235: ...diginp2 X01 0 digital input 2 not active actual diginp3 X02 0 digital input 3 not active 1 Select and confirm the condition line of the appropriate parameter A selection window opens 2 Select and conf...

Страница 236: ...ower failure not active test active X56 0 Analog test was not started cal active X57 0 For internal use only tare active X58 0 Instrument is not tared marker bit 1 X64 0 Marker bit 1 not set after pow...

Страница 237: ...r bit 1 set after power on the markers are set to 0 marker bit 2 X65 1 Marker bit 2 set after power on the markers are set to 0 marker bit 3 X66 1 Marker bit 3 set after power on the markers are set t...

Страница 238: ...dow opens 2 Select below zero and confirm 3 Select Output 2 and confirm Output 2 Output 2 remains active as long as the weight is not above Max X33 0 A selection window opens 4 Select above MAX and co...

Страница 239: ...i o parameter Note This menu item is only available if under Operating parameter Application Standard has been selected Output 3 Output 3 is true active when the weight is zero d X36 1 6 Select center...

Страница 240: ...analog output under Analog output parameter The analog output can be configured according to the table below Parameter table Menu item Selection Description no output Analog output is unused Gross D08...

Страница 241: ...t output to 20 mA Output if Max linear Output increases above 20 mA until the limit is re ached Weight at 0 4 mA Weight value for 0 4 mA output Weight at 20 mA Weight value for 20 mA output Quit the m...

Страница 242: ...date PRxxxx Application Version number and application creation date BIOS Version number and BIOS creation date Board number Nine digit board number This line shows the following information 2 Press...

Страница 243: ...tery Status display ok voltage OK low voltage too low CAL switch Status display opened opened no write protection closed closed write protection is active This line shows the following device statuses...

Страница 244: ...puts 4th line Standard interface digital I Os 5th line In this case Slot 4 ProfiBus DP interface The tool tip displays the version number 6th line Standard interface weighing electronics The tool tip...

Страница 245: ...e number Last entry Date Date of the last entry Time Time of the last entry Gross weight Weight type of the last entry Select the next or previous entry Note For more information see Chapter 9 2 7 The...

Страница 246: ...orrection The zero point correction in use is displayed Communication error count communication error counter The communication errors time frames exceeded for the load cells are counted here in ascen...

Страница 247: ...n Weight value is within tolerances Orange Weight value is above maximum capacity Emax max capacity of load cell No ID The Serial number is hidden Show ID The Serial number is visible 2 Select the des...

Страница 248: ...number the higher the probabi lity of a faulty load cell Temperature Current measured temperature Max temperature Max measured temperature Min temperature Min measured temperature Max weight value at...

Страница 249: ...e front panel keys 8 2 1 Configuring production mode using the front panel keys You can select between the following production modes Net filling B1 Net Discharge B4 Note See also Chapter 8 4 1 1 Pres...

Страница 250: ...See also Chapter 8 4 2 Configuring digital inputs 1 Press SEtuP appears on the display 2 Press OK The Cd calibration menu flashes Cd 000 appears on the display 3 Press multiple times to select IM 151...

Страница 251: ...y 10 Press the cursor keys to enter the desired SPM address see Chapter 13 4 11 Press OK dI 176 appears on the display 12 Press OK InP 3 appears on the display 13 Press OK 0001 appears on the display...

Страница 252: ...ppears on the display 24 Press OK 0001 appears on the display 25 Press the cursor keys to enter the desired SPM address see Chapter 13 4 26 Press OK do 172 appears on the display 27 Press OK out 3 app...

Страница 253: ...splay 10 Press to select PC 161 11 Press OK SEQnr sequence number appears on the display 12 Press OK 0000 appears on the display 13 Press the cursor keys to enter the desired number 14 Press OK SAVE a...

Страница 254: ...rmfeed SEq sequence number dAtE date and time SEtP set point tArE tare weight nL CR LF carriage return and line feed nEt appears on the display 25 Press the cursor keys to select the desired option 26...

Страница 255: ...r 7 17 3 Procedure 8 4 Configuration via a notebook PC 8 4 1 Configuring production mode The following modes are configured under the Configuration mode menu item Filling mode Interaction mode Press t...

Страница 256: ...2 Select Dosing mode and confirm A selection window opens 3 Select the desired filling mode see Chapters 8 4 1 1 and 8 4 1 2 and confirm 4 Select Interaction mode and confirm X3 Process Indicator PR...

Страница 257: ...cally added coarse fine A fixed overshoot value is configured Net gross tare 5 Select the desired interaction mode see Chapter 8 4 1 3 and confirm 6 Press to exit the menu A prompt window opens 7 Pres...

Страница 258: ...point overshoot is reached Calming Time to wait during which the overshoot is effective and scale vibrations may settle Tolerance checking The weight is determined and checked against the tolerance v...

Страница 259: ...nce values 8 4 1 3 Interaction mode You can choose between the following control operating modes of the device in production Remote proc control via OPC ModBus and or fieldbus VNC Virtual Network Comp...

Страница 260: ...anged after a restart 1 Select Configuration digital IOs and confirm 2 Select Configuration digital inputs and confirm The following window opens 3 Select inputs 1 3 Use the keypad to enter and confir...

Страница 261: ...gured under the Configuration material menu item 4 Press to exit the window and to save the changes 5 Select Configuration digital outputs and confirm 6 Select outputs 1 3 Use the keypad to enter and...

Страница 262: ...ed material do not match the parameters of the current calibration The configuration window appears 2 Press the OK softkey 3 Press the Default softkey All values are reset 4 Enter the material name an...

Страница 263: ...flow Overshoot OVS Input Material overshoot tolerance Input Tolerance above below set point Calming time Input Calming time 5 Configure additional materials if necessary 6 Use the Print all or Print...

Страница 264: ...he configuration window appears 2 Select Number printouts enter the number of printouts between 1 10 and confirm 3 Select Sequence number change the sequence number if necessary and confirm 4 Select U...

Страница 265: ...Material ID 2 Selection none no printout selected if fewer than 6 elements will be printed Material name Gross weight Net weight Tare weight Set point Date Time printed in format DD MM YYYY HH MM SS...

Страница 266: ...softkey Press the Stop softkey to stop the process You can then press the Restart softkey to restart the process The material product is filled 4 Once the set point is reached the Start softkey can be...

Страница 267: ...goes back to the weight display Operate all display elements 1 Press and hold and also press A number is displayed for each key The number increases by 1 each time a key is pressed 0 0 0 0 0 0 2 For...

Страница 268: ...nitor Test mode Internal Test mode External 1 Plug the test connector see Chapter 18 2 into the RS 232 interface 2 Select the interface and confirm passed ok failed no data error The results are displ...

Страница 269: ...isplayed display The physical inputs and outputs of the system plant are deactivated and passive in secured condition External Deactivated PLC The physical inputs are displayed dis play Output values...

Страница 270: ...the Adjust softkey The window for the 1st value opens 3 Enter and confirm e g the value for 4 mA measured by the connected PLC under Measured The window for the 2nd value opens X3 Process Indicator PR...

Страница 271: ...e for 20 mA measured by the connected PLC under Measured A prompt window opens 5 Press the Yes softkey to save the settings The following window opens 6 Press the Reset softkey to reset to the factory...

Страница 272: ...s A prompt window opens 7 Press the Yes softkey to reset to the factory settings 8 Press to return to the previous window 1 Select and confirm the appropriate line The Monitor test mode is active The...

Страница 273: ...he functionality of the PLC application see Chapter 9 1 4 An info window is shown briefly The Internal test mode is active 4 Press to return to the previous window 5 Press the Stop PLC softkey 6 Enter...

Страница 274: ...pter 9 2 2 Configuration printout Display configuration printout saving and printing out as a text file see Chapter 9 2 3 Log files Display logfiles saving and printing out as a text file see Chapter...

Страница 275: ...Click on the Indicator Pop up Window menu item in the WEB menu A separate status window opens in which the weight of the weighing point is displayed with unit and the status symbols 4 Click on the sym...

Страница 276: ...4 Create and open the required directory e g on the notebook 5 Click on Save to save the text file in the relevant directory 6 Press the right mouse button 7 Click on Print 8 Select the connected prin...

Страница 277: ...e and open the required directory e g on the notebook 5 Click on Save to save the graphic file in the relevant directory 6 Press the right mouse button 7 Click on Print 8 Select the connected printer...

Страница 278: ...Always A Weight type G gross N net T tare PT preset tare Weight with unit Depending on the configuration the gross and or net and or tare weight are saved using the front panel keys oP 063 see Chapter...

Страница 279: ...bottom of the screen 2 Click on the icon Commas are used in the CSV file to separate the individual elements of an entry The file automatically opens e g in Microsoft Excel 3 Carry out a conversion i...

Страница 280: ...eight 4 Read request D18 Tare weight 5 Read request D19 Sequence number for alibi memory 6 Read request D21 Date in BCD display 7 Read request D22 Time in BCD display 9 2 7 3 Exporting alibi data reco...

Страница 281: ...itch is open The IP address and the Hostname remain unaffected Click the Backup of Earom menu item in the WEB menu The backup restore menu is shown on the display 1 Click on Backup to create a backup...

Страница 282: ...on the display 3 Press OK FLASH follows on the display bIoS appears on the display 4 Press 2 dEL EAr appears on the display 5 Press OK ErAS E and then no appear on the display 6 Press YES appears on t...

Страница 283: ...his allows for the parameters to be re entered manually in the device A backup of version 4 00 cannot be loaded onto a device with version 4 00 See Chapters 9 2 3 and 9 4 9 5 1 Updating in network wit...

Страница 284: ...n 1 Press SEtuP appears on the display 2 Press bIoS appears on the display 3 Press OK FLASH follows on the display bIoS appears on the display 4 Press OK EtHEr appears on the display 5 Press OK The de...

Страница 285: ...program is started 7 Make sure to note down the digit sequence because it must be entered into the FlashIt 32 window subsequently Now the device is ready to load the software FLASH appears on the disp...

Страница 286: ...display 4 Press OK EtHEr appears on the display 5 Press OK The device waits for address assignment by the server 172 024 The more significant part of the IP address is displayed for approx 2 seconds...

Страница 287: ...1 Notebook PC IP address 192 24 22 2 Device and notebook PC have the subnet mask 255 255 255 0 The FlashIt 32 program in a directory on the enclosed CD ROM is installed on the notebook PC The FlashIt...

Страница 288: ...e device while it is switched on may have life threatening consequences Product operation must be performed by trained and qualified personnel who are aware of and able to deal with the related hazard...

Страница 289: ...ds dHCP appears on the display 6 Make sure to note down the digit sequence because it must be entered into the FlashIt 32 window subsequently 021 081 The less significant part of the IP address is dis...

Страница 290: ...nto the Select window or use Copy and paste 2 Under COM Port select Network and enter the devices IP addresses 3 Click Flash to start the procedure As soon as software loading is completed the device...

Страница 291: ...t tared Tare weight 0 if not tared Read data Address Read Word display in binary format W201 0E00000T 00000000 E ADC error T Instrument is tared W203 000S00MZ 00000000 S Device is in setup mode M Weig...

Страница 292: ...Bit address Value in 32 bit integer format W101 Tare device Write value 256 Reset the tare of the device Write value 512 X3 Process Indicator PR 5410 10 ModBus protocol Minebea Intec EN 290...

Страница 293: ...and 1 stop bit The commands to the transmitter are printable ASCII characters starting with LF 0A hex and ending with CR 0D hex The transmitter sends a reply on each received command after approx 100...

Страница 294: ...standard application The interface works with an 8 byte write window and an 8 byte read window The fieldbus exchanges its data cyclically with each slave That means During each cycle 8 bytes are writ...

Страница 295: ...te data MSB Byte 2 Write data MSB Byte 3 Write data LSB Byte 4 Read_Value_Select Byte 5 Write_Value_Select Byte 6 Direct control bits Byte 7 Direct control bits The first four bytes are used to write...

Страница 296: ...o write data in the PR 5410 e g a limit value is set to 100 kg The master can also read weight values or other data from the PR 5410 The write and read windows are always required for this Safe data e...

Страница 297: ...the device with a Set Bit After execution of the operation the bit should be reset Variable Function Write_Value The weight value is transmitted as a binary 32 bit value with plus or minus sign Type...

Страница 298: ...Read_Value The weight value is transmitted as a binary 32 bit value with plus or minus sign Type of data DINT Read_Value_Selected Acknowledgment of the transmitted value Write_Active The function sele...

Страница 299: ...lue_Select Read_Value_Selected When the master has to read from the device the register number is transmitted to Read_Value_Select in the input area The result is specified with Read_Value_Selected in...

Страница 300: ...tive has been set Write 0 to Write_Value_Select Reset the Write_Active bit 12 2 4 4 Reading bit Single bits can be read only by reading a register The procedure is described in Chapter 12 2 4 3 12 2 4...

Страница 301: ...ng etc over the fieldbus Example Reading the gross weight The master writes value 8 in Read_Value_Select byte 4 of the input area Input section Byte Name Description 0 1 2 3 4 8 Gross 5 6 7 Bit 7 Bit...

Страница 302: ...pe and software version see table below Standard sequence Sequence for DeviceNet and EtherNet IP Byte 0 TYPE MSB Byte 0 SUBVERSION Byte 1 TYPE LSB Byte 1 MAIN VERSION Byte 2 MAIN VERSION Byte 2 TYPE L...

Страница 303: ...r 3 E1 Arithmetic error overflow Error 1 E9 No communication with xBPI scale Error 9 CmdError Error during execution cmdError e g the taring operation is not handled because the scale is not stable Th...

Страница 304: ...0 EXPO Byte 1 UNIT Byte 2 STEP Byte 3 LASTER ROR Bit Description EXPO One byte for the position of the decimal point content in decimal form 0 255 0 000000 1 00000 0 2 0000 00 3 000 000 4 00 0000 5 0...

Страница 305: ...mber Read Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0 Byte 0 board number MSB Byte 1 board number MSB Byte 2 board number MSB Byte 3 board number LSB e g 148388723 08D83B73hex 12 2 6 8 Register 7...

Страница 306: ...log output value num 0 20000 corresponds to 20 mA Register 31 Fixed value for preset tare see also SetFixTare GetFixTare in Chap ter 12 2 6 3 12 2 6 12 Registers 80 92 State controlled action bits Wri...

Страница 307: ...nsition controlled for writing once The bit is set as Write_Value_Select with the specified number see Chapter 12 2 4 3 Register 112 SetZero Register 113 SetTare Register 114 ResetTare Register 115 Se...

Страница 308: ...Filling parameters are written at each start by the PLC B Filling parameters are stored in the device PR 5410 Requirements The system is ready The device has been checked to ensure it is ready for th...

Страница 309: ...Hold_state bit is active Example Set cancel bit with 0001 0000 in byte 1 Filling starts again with the rising edge and the system leaves the hold status hold state and goes back to the ready status r...

Страница 310: ...reserved Bit 7 reserved BOOL Bit 0 EasyFill adopts the sent material parame ters from byte 4 with the rising edge of this bit A valid set point byte 4 7 that can be coded as DINT or REAL is required...

Страница 311: ...switching from coarse flow to fine flow Byte 12 15 Material overshoot OVS DINT REAL Weight Byte 16 19 Calming time DINT ms Byte 20 23 Tolerance above set point DINT REAL Weight Byte 24 27 Tolerance be...

Страница 312: ...possible within the menu items Bit 7 status controlled 0 fill 1 empty Byte 3 Bit 0 error in the application Bit 1 filling is active Bit 2 system ready Bit 3 filling stopped Bit 4 coarse flow Bit 5 fi...

Страница 313: ...ew result Byte 48 51 Date of filling result DINT Example 20161116h corresponds to November 16 2016 Byte 52 55 Time of filling result DINT Example 14153199h corresponds to 14 15 31 and 99 ms Byte 56 59...

Страница 314: ...and can be reset by bit 0 in byte 2 Bit 6 Filling is not possible within the menu items Bit 7 status controlled 0 fill 1 empty Byte 3 Bit 0 error in the application Bit 1 filling is active Bit 2 syst...

Страница 315: ...overshoot OVS DINT REAL Current set point Byte 48 51 Calming time DINT Current set point Byte 52 55 Tolerance above set point DINT REAL Current set point Byte 56 59 Tolerance below set point DINT REAL...

Страница 316: ...the BOOL data type are executed with a rising edge Data type Description Value range BOOL bool 0 FALSE or 1 TRUE SINT short integer 128 to 127 INT integer 32 768 to 32767 DINT double integer 231 to 23...

Страница 317: ...R W Function X0 X2 BOOL R Digital input 1 3 X8 10 BOOL R Digital output 1 3 X16 18 BOOL R Standard only Output limit 1 3 B4 X32 X33 X34 X35 X36 X37 X38 X39 BYTE BOOL BOOL BOOL BOOL BOOL BOOL BOOL BOOL...

Страница 318: ...he device X115 BOOL W Start the test mode X116 BOOL W Finish the test mode X117 BOOL W Reset the power fail signal X118 BOOL W Set fixed tare weight value as tare X119 BOOL W Store the current gross w...

Страница 319: ...e if no BCD card is installed Standard only BCD value for transparent mode D21 DINT R Report Date D22 DINT R Report Time BCD display D23 DINT R Activity counter test of communication with device D24 D...

Страница 320: ...command 2 Start ID 7 Start ID 8 Start ID 9 Start ID 10 Print report Reset error status D67 0 B148 X1182 X1183 BYTE BOOL BOOL W W W EasyFill only Batching command 3 Load selected material Save selected...

Страница 321: ...3 REAL R Report Net weight as floating point number R274 REAL R Report Tare weight as floating point number R280 REAL R Standard only Limit Limit 1 on as floating point number R281 REAL R Standard onl...

Страница 322: ...ample Production start Write material parameters D44 D49 Start batching X1168 Monitor status B144 B145 Once batching is complete read report data X3 Process Indicator PR 5410 13 SPM Minebea Intec EN 3...

Страница 323: ...fter initial start up the date and time must be checked and set if necessary under Date Time 14 1 2 Changing the battery for date time The device is equipped with a lithium battery for backing up the...

Страница 324: ...use a cleaning agent suitable for that particular working environment Use soft sponges brushes and cloths Spraying with water or blasting with compressed air is not permitted 1 Unplug device from mai...

Страница 325: ...unsorted municipal waste so that it can then be recycled Before disposing of or scrapping the product any batteries should be removed and taken to a suitable collection point Please see our T Cs for f...

Страница 326: ...ct Error 3 Error 4 Value exceeds display The weight value is not displayed Too many digits have been set Error 4 Error 6 No sense voltage No sense voltage Load cells not connected Sense line or supply...

Страница 327: ...s Serial ports parameter xBPI port Slot 1 2 RS 485 is selected Error Error 12 Scale not ready The scale is not ready for weighing The device is in the warm up phase The device is in automatic taring m...

Страница 328: ...ommunication Error Error 13 Wrong configuration The number of load cells does not match the con figuration Error Error 15 Wrong serial number Serial number of scale does not match the num ber set in t...

Страница 329: ...e is displayed when the dead load entered in mV V plus scale interval in mV V is higher than 3 mV V 36 mV This message displays if the scale is not stable Remedy Check the mechanical function of the s...

Страница 330: ...e Err 21 BIOS is too old for the firmware Err 22 Firmware is too old for the BIOS Err 23 Boot not possible ROM error No firmware found Err 25 Plug in card combination is not permissible Err 29 Flash p...

Страница 331: ...the weighing point is interrupted Err 72 This message is displayed for internal weighing point during the warm up phase when the device is in the automatic taring mode if the device has been switched...

Страница 332: ...unication is not assigned 16 6 Error numbers LAST_ERROR 16 6 1 Weighing point error Number Display Cause 2 in use Weighing point is in use e g by an application 6 test active Test is active no weights...

Страница 333: ...umber is invalid 13 Invalid fieldbus com mand Fieldbus action is invalid e g simultaneous start and stop com mand 15 Cannot read from earom Error when reading a material entry from the hard drive memo...

Страница 334: ...th 252 April 2010 Current number Is increment according to the year group table of Minebea Intec 17 3 General data The following characteristics are valid after a warm up time of at least 60 minutes r...

Страница 335: ...1000 MHz EN 61000 4 3 10 V m Housing Electrostatic discharge ESD EN 61000 4 2 6 8 kV Fast transients burst EN 61000 4 4 1 kV Peak voltages surge 1 2 50 s EN 61000 4 5 1 2 kV Signal and control lines...

Страница 336: ...igital filter Can be activated active 4th order low pass Characteristic Bessel aperiodic Butterworth Tscheby scheff Limit frequency configurable 17 5 3 Accuracy and stability Accuracy 0 5 V e corr to...

Страница 337: ...d cell cable Length of the connecting cable between weighing platform and instrument Cable type LiYCY max 500 m 17 6 Mechanics 17 6 1 Design Metal housing made of aluminum with steel cover and rear pa...

Страница 338: ...screws 5312 505 18015 Knurled screw set for brackets 5322 214 90581 Display PCB with screen 5312 253 98006 Fuse 1 6 A T 250 V 5 20 mm for 24 V de vices 5312 455 98052 Keypad overlay 5322 214 90084 Mai...

Страница 339: ...18 2 Test connector for the RS 232 interface Pin assignment 18 Appendix X3 Process Indicator PR 5410 EN 337 Minebea Intec...

Страница 340: ...Published by Minebea Intec GmbH Meiendorfer Strasse 205 A 22145 Hamburg Germany Phone 49 40 67960 303 Email info minebea intec com www minebea intec com...

Отзывы: