16.6.2
Error in the "EasyFill" application
Number
Display
Cause
0
No error.
1
fatal error
Weight error; weighing point is faulty.
2
Material ID invalid
Number <0 or >10 entered.
3
Material name inva-
lid
Material name is invalid.
4
Set point invalid
Set point + gross value > scale end value, (B1 mode) or gross value.
5
Preset point invalid
Value > set point.
6
OVS invalid
Value > set point.
8
Neg. tol. invalid
Value > set point.
9
Pos. tol. invalid
Value > set point.
10
Sequence number
invalid
Sequence number is invalid.
13
Invalid fieldbus com-
mand
Fieldbus action is invalid (e. g., simultaneous start and stop com-
mand).
15
Cannot read from
earom
Error when reading a material entry from the hard drive memo-
ry (EAROM) → hardware error
16
Cannot write to ea-
rom
Error when writing a material entry to the hard drive memory (EA-
ROM) → hardware error
17
Action not allowed
Fieldbus action is not permitted.
Example: Starting filling during an ongoing filling process or star-
ting filling when querying the system setup.
18
Weight unit of mate-
rial invalid
The unit of the weighing point does not match the unit of the ma-
terial.
16 Error messages
X3 Process Indicator PR 5410
EN-331
Minebea Intec