7.5.6
Search for Pendeo load cells and set dead load using the front-panel keys
In this menu, a search for connected load cells is conducted and the dead load is set.
The load cell type (Pendeo Truck or Pendeo Process) is automatically detected.
Note:
Assigning the Pendeo load cells is only possible with a notebook/PC; see Chapter
.
In the process, the dead load is deleted and a recalibration must be carried out.
Procedure
The "Cd" calibration menu flashes.
Cd 000
appears on the display.
3.
Press
to change the parameter number ("000" flashes).
4. Press
multiple times to select parameter number "003."
5.
Press OK.
dEAdLo
appears on the display.
6. Press OK.
LoAd
appears on the display.
7.
Press
to select "MvoLt."
8. Press OK.
The mV/V value for the dead load appears on the display.
9. Note the displayed value.
10. Press OK.
Cd 004
appears on the display.
11. Press OK.
SPAn
appears on the display.
12. Press OK.
MvoLt
appears on the display.
13. Press OK.
The mV/V value for the maximum load appears on the display.
14. Note the displayed value.
15. Press "Exit" to exit the Setup menu.
1.
Press
.
SEtuP
appears on the display.
X3 Process Indicator PR 5410
7 Commissioning
Minebea Intec
EN-134