The weight unit for the CAL weight may differ from the unit in the device. Conversion
is automatic.
Note:
Error messages, calibration see Chapter
.
7.14.8
Calibrating with mV/V
The scale can be calibrated without weights. During input of the load cell mV/V value, the
acceleration of gravity at the place of installation can be taken into account.
The STAR load cell data is based on the acceleration of gravity in Hamburg, Germany:
9.81379 m/s
2
.
1.
Press the [by load] softkey.
2.
Place the CAL weight on the scale.
3.
Enter the weight of the CAL weight.
4. Confirm the entries.
5.
Press
to select the weight unit.
6. Confirm entries with
or
.
Weight value, weight unit and measuring signal in mV/V corresponding to this
value are displayed in the line [Calibrated at].
The verification is displayed by "Setting Span by load…".
X3 Process Indicator PR 5410