IBP Monitoring
15-13
Prompt Information Related to Calibration
Take channel 1 as an example. The monitor may give the following prompt
information after the calibration has been finished.
CH1 SUCCESSFUL CALIBRATE
Channel 1 works normally and can be applied to IBP monitoring.
CH1 SENSOR OFF, FAIL
Verify the transducer of channel 1 is connected to the monitor, and then perform the
calibration again. If the monitor continues to give this prompt information, contact
our Customer Service.
IN DEMO, FAIL
Make sure the monitor is not in DEMO mode, and perform the calibration again. If
the monitor continues to give this prompt information, contact our Customer
Service.
PRESSURE OVER RANGE, FAIL
Verify the preset calibration value is reasonable, and perform the calibration again. If
the prompt information is still displayed, contact our Customer Service.
PULSATILE PRESSURE, FAIL
Verify the pressure value indicated by the sphygmomanometer keeps unchanged,
and then perform the calibration again. If the prompt information is still displayed,
contact our Customer Service.
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...