Monitoring
17-23
17.4.2 Preparations for CO
2
Measurement
1. Plug the water trap into its receptacle before the measurement.
2. Power on the monitor and open the CO2 SETUP menu. In the menu, set the
WORK MODE to MEASURE.
3. The "CO2 WARM UP" technical information is displayed on the screen until
the sensor reaches its work temperature.
Figure 17-12 CO
2
Module Airway Connections
When the monitor is powered on for the first time, the CO
2
module enters the
STANDBY mode by default. To activat the CO
2
module, you must change the work
mode to MEASURE. The work mode of the CO
2
module keeps unchanged when the
monitor is restarted. For instance, if the monitor is power off when the CO
2
module
is working in the MEASURE mode, the CO
2
module will automatically enter the
MEASURE mode after the monitor is restarted. For details about WORK MODE,
please refer to the section
Other Setup
of
17.4.3 CO2 Setup Menu
.
Mainstream
Sensor
Airway
Adapter
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...