CO
2
Monitoring
17-7
AWRR ALM HI
Upper alarm limit of AwRR
AWRR ALM LO
Lower alarm limit of AwRR
APNEA ALM
Determines the apnea alarm delay.
If the apnea of the patient exceeds the preset apnea alarm
delay, the monitor triggers an alarm and gives the " CO2
APNEA " alarm message.
Options: 10S, 15S, 20S, 25S, 30S, 35S and 40S.
SWEEP
Waveform speed
Options: 6.25, 12.5 and 25.0.
Unit: mm/s.
UNIT
Options: mmHg, kPa and %.
WAVE SCALE
Options: HIGH and LOW.
This option allows you to adjust the amplitude of the CO
2
waveform.
WORK MODE
Options: MEASURE and STANDBY.
To start the CO
2
monitoring, select the MEASURE mode.
In the STANDBY mode, the air pump and infrared ray
source of the CO
2
module are disabled. Hence, decreases
the power consumption and prolongs the operating life of
infrared ray source and the whole CO
2
module.
NOTE
z
The apnea alarm cannot be disabled.
z
When the CO
2
monitoring is not required, it is recommended the water
trap not be connected and the work mode be set to STANDBY.
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...