A.3
Power Source Specifications
AC mains
Input voltage
100 to 240V
Frequency 50/60Hz
Power 140VA
Fuse T
3A
Internal battery
Number of batteries
2
Type
Sealed lead-acid battery or lithium-ion battery
Time to shutdown
5 to 15min (after the first low-power alarm)
Sealed lead-acid battery
Nominal voltage
12VDC
Capacity 2.3Ah
Operating time
48 minutes or 120 minutes typical when powered by one or two
new fully-charged batteries respectively (25
℃
, ECG, SpO2,
NIBP measurement per 15 minutes).
Charge time
A maximum of 6 h for each battery, and a maximum of 12h for
both (in the running status or standby mode)
Lithium battery
Rated voltage
11.1VDC
Capacity
4.4Ah
Operating time
120 minutes or 300 minutes typical when powered by one or two
new fully-charged batteries respectively (25
℃
, ECG, SpO2,
NIBP measurement per 15 minutes).
Charge time
A maximum of 6.5h (in the running status or standby mode)
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...