ECG/RESP Monitoring
11-40
ALM REC
ST alarm recording
ON: The monitor starts recording when an ST alarm occurs;
OFF: The monitor does not record when an ST alarm occurs.
ANT ALM HI
Determines the upper limit of the ST alarm in the ANT group;
2.0mV is the highest.
ANT ALM LO Determines the lower limit of the ST alarm in the ANT group;
-2.0 mV is the lowest.
INF ALM HI
Determines the upper limit of the ST alarm in the INF group;
2.0mV is the highest.
INF ALM LO
Determines the lower limit of the ST alarm in the INF group;
-2.0 mV is the lowest.
LAT ALM HI
Determines the upper limit of the ST alarm in the LAT group;
2.0mV is the highest.
LAT ALM LO
Determines the lower limit of the ST alarm in the LAT group;
-2.0 mV is the lowest.
12-lead review
Select 12 LEAD RECALL >> In ECG SETUP menu. The following menu appears.
Figure 11-28 12-lead review
In this menu, you can recall 80 12-lead analysis results. When the number of 12-lead
results exceeds 80, the earliest results will be deleted.
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...