Alarms
6-8
6.3.5
Status Switchover
1. In the Normal status,
Press the SILENCE key for less than 2 seconds to switch to the Alarms Paused
status, or
Press the SILENCE key for 2 seconds or more to switch to the System Silenced
status.
2. In the Alarms Paused status,
Press the SILENCE key for less than 2 seconds to switch to the Normal status,
or
Press the SILENCE key for 2 seconds or more to switch to the System Silenced
status.
3. In the System Silenced status,
Press the SILENCE key for less than 2 seconds to switch to the Alarms Paused
status, or
Press the SILENCE key for 2 seconds or more to switch to the Normal status.
4. In the Alarms Silenced status,
Enter the ALARM SETUP menu and set ALARM VOL to any setting other
than 0. The alarm sound is restored.
Enter the USER MAINTAIN menu and set ALARM SOUND OFF to
DISABLE. The system restores the default alarm volume.
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...