CO
2
Monitoring
17-20
Calibration Procedure
1. Verify the module enters the Full Accuracy Mode.
2. Connect a gas bottle with the CO
2
sensor connector on the monitor as shown
below.
Figure 17-11 Connections for Calibration
3. Fill the gas bottle with a standard gas of certain CO
2
content (4% to 6%), and
input the gas to the monitor.
4. Open the CO2 USER MAINTAIN menu, and set the CO
2
field to a value,
which is the same with the CO
2
content of the gas bottle.
5. When the CO2 USER MAINTAIN menu displays the currently measured CO
2
content, wait until a stable reading is indicated, and then select the CONFIRM
CAL option to start calibrating the CO
2
module.
6. If the calibration succeeds, “CALIBRATE SUCCESS!” will be displayed in the
CO2 USER MAINTAIN menu. If fails, “CAL FAIL, TRY AGAIN!” will be
displayed, and re-calibration is required.
7. Select the EXIT option to exit the CO2 USER MAINTAIN menu.
Monitor
Gas bottle
T-shape connector opening
to the atmosphere
Decompression
valve
Tube
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...