9.1
Overview
The monitor is able to store important patient data so that the user can review and
record the data as desired.
Trend Graph Recall
You can review the latest 1-hour trend graph of a measured parameter displayed
every 1 or 5 seconds, or the latest 96-hour trend graph displayed every 1, 5 or 10
minutes.
Trend Table Recall
You can review the latest 96-hour trend table data of a measured parameter.
NIBP Recall
You can review 800 NIBP measurement results, each of which include a systolic
pressure, a mean pressure, a diastolic pressure , a pulse rate and a measurement time.
Alarm Event Recall
You can review the latest 70 parameter alarm events, as well as the 8, 16, or
32-second waveforms stored at the time of the alarm.
Arrhythmia Event Recall
You can review the latest 80 arrhythmia events and the related 8-second waveforms.
This chapter only gives introduction to the first four recalls mentioned above. For
details about the arrhythmia event recall, please refer to
11 ECG/RESP Monitoring
.
NOTE
z
When your monitor is powered off or the power is interrupted
accidentally, all the stored data is lost, if the monitor is not equipped
with the fuction of “Power-off Data Storage”.
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...