Appendices
20-20
A.16 AG
Specifications
Measurement technique
Infrared absorption
Measurement mode
Side stream
AG functions
Meets requirements of ISO9918, ISO11196, EN12598 and
ISO7767
Warm-up time
45 seconds (warming-up status)
10 minutes (ready-to-measure status)
Adult/Pediatric
120, 150, 200 ml/minute (user-selectable)
Sampling flow
(sidestream)
Neonatal
70, 90, 120 ml/minute (user-selectable)
Gas type
CO
2
, N
2
O, O
2
(optional), Des, Iso, Enf, Sev, Hal
Measurement range
CO
2
:
N
2
O:
Des:
Sev:
Enf, Iso, Hal:
O
2
:
AwRR:
0 to 30%
0 to 105%
0 to 30%
0 to 30%
0 to 30%
0 to 105%
2 to 100 BrPM
Resolution
CO
2
:
AwRR:
1 mmHg
1 BrPM
Gas
Range (%REL)
Precision (%ABS)
0 to 1
±
0.1
1 to 5
±
0.2
5 to 7
±
0.3
7 to 10
±
0.5
CO2
> 10
Not specified
0 to 20
±
2
N2O
20 to 100
±
3
0 to 1
±
0.15
1to 5
±
0.2
5 to 10
±
0.4
Precision
Des
10 to 15
±
0.6
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...