System Menu
4-16
4.4.5
Data Output
Select DATA OUTPUT>> in SYSTEM SETUP menu. The following menu appears.
Figure 4-11 Data Output
Output Procedure
1. Disconnect all patient cables connected to the monitor.
2. Verify the monitor is connected to the PC and the PC is running the Patient
Information Recall System software.
3. Select from the five data source options: TREND, PARAM ALARM
(parameter alarm), DRUG CALC (drug calculation), NIBP and ARR ALARM
(arrhythmia alarm).
4. Select OUTPUT in the menu and the prompt ”CONNECTING…” is shown
aside. If you exit the DATA OUTPUT menu at this time, the prompt will be
displayed in prompt information area at the lower left corner of the screen.
5. If the connection is available, the data will be output to the PC. For more
information, please refer to the help information of the Patient Information
Recall System software.
NOTE
z
If no data source is selected or the previous data output has not
finished, the OUTPUT option in the DATA OUTPUT menu will be inactive.
z
During data output, the NEW PATIENT option in the PATIENT SETUP
menu is inactive.
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...