Appendices
20-13
A.11 NIBP
Specifications
Measurement technique
Auto oscillation
Displayed parameters
Systolic pressure, diastolic pressure and mean pressure
Mode of operation
Manual, auto and continuous
Measurement interval in
auto mode
1/2/3/4/5/10/15/30/60/90/120/180/240/480 minutes
Measurement time in
continuous mode
5 minutes
mmHg Adult Pediatric
Neonate
Systolic
pressure
40 to 270
40 to 200
40 to 135
Diastolic
pressure
10 to 210
10 to 150
10 to 100
Measurement range in
normal mode
Mean pressure
20 to 230
20 to 165
20 to 110
Measurement precision
Maximum average error:
±
5mmHg
Maximum standard deviation: 8mmHg
Resolution 1mmHg
Over-pressure protection
Adult:
Pediatric:
Neonate:
297±3 mmHg
240±3 mmHg
147±3 mmHg