CO
2
Monitoring
17-18
NOTE
z
When measuring a moist gas of saturated water vapor at the body
temperature and the ambient pressure, turn ON BTPS; when measuring
a dry gas at the ambient temperature and pressure, turn OFF BTPS.
AUTO
STANDBY
Range: 0 to 60min.
If no respiration waveform is detected in the selected time,
the CO
2
module enters the standby mode automatically.
When AUTO STANDBY is set to 0min, it means the CO
2
module is in MEASURE state. If the sampling line is not
connected, the CO
2
module enters the standby mode
automatically 3 minutes later.
ZERO CAL
This option allows you to zero manually so as to eliminate
the adverse affect of baseline drift in the process of
measurement.
DEFAULT >>
Select DEFAULT >> to access the CO2 DEFAULT
CONFIG menu. You can select either FACTORY
DEFAULT CONF or USER DEFAULT CONF. After
finishing the selection and exiting the menu, a dialog pops
up asking for confirmation of your selection.
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...