Appendices
20-40
C.3 Prompt
Messages
Prompt messages
Cause
Measure
ECG1 SIGNAL
SATURATION
ECG2 SIGNAL
SATURATION
Signals of abrupt change interfere with the
ECG signal.
Check whether the electrodes and
leads are well connected.
SEARCH PULSE
The SpO
2
module is searching the pulse.
Wait till the end of the searching.
Welch Allyn CO2 module
CO2 STANDBY
The CO
2
module enters the energy-saving
status when switching from the Normal mode
to the Standby mode.
None
CO2 WARM UP
The CO
2
module is starting and warming
itself up.
Wait for the CO
2
module to finish
warming itself up.
CO2 SENSOR START
UP
The CO
2
sensor is starting.
Wait for the CO
2
sensor to finish
the startup.
Mindray CO2 module
CO2 STANDBY
The CO
2
module enters the energy-saving
status when switching from the Normal mode
to the Standby mode.
None
CO2 START UP
The CO
2
module is starting.
Wait for the CO
2
module to finish
the startup.
CO2 CALIBRATE
ZERO
The CO
2
module is in the zeroing status.
Wait for the CO
2
module to finish
the zeroing.
CO2 WARM UP
The CO
2
module is warming itself up after
startup.
Wait for the CO
2
module to finish
warming itself up.
Oridion CO
2
module
CO2 STANDBY
The CO
2
module enters the energy-saving
status when switching from the Normal mode
to the Standby mode.
None
CO2 START UP
The CO
2
module is starting.
Wait for the CO
2
module to finish
the startup.
CO2 CALIBRATE
ZERO
The CO
2
module is in the zeroing status.
Wait for the CO
2
module to finish
the zeroing.
CO2 SENSOR START
The CO
2
sensor is warming itself up after
Wait for the CO
2
module to finish
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...