ECG/RESP Monitoring
11-43
11.7 RESP Monitoring
11.7.1 Overview
Respiration is detected by measuring thoracic impedance. The monitor measures the
change of the impedanec between the RA and LA electrodes of the ECG lead I, or
the RA and LL electrodes of the ECG lead II, and produces a respiration waveform
as shown below.
Figure 11-31 Respiration Waveform and Parameter
1. Waveform
name.
2. RESP lead: I or II
3. Waveform gain: 7 options are available: ×0.25, ×0.5, ×1, ×2, ×3, ×4, ×5. If the
gain is too large, the upper part of the waveform may not be normally
displayed.
4. RESP label: Selecting this label, you can open the RESP SETUP menu.
5. RR: Respiration rate numerics.
NOTE
z
Respiration monitoring is not recommended on patients who are very
active, as this will cause false alarms.
z
It is recommended to set Waveform gain to 1 when external
electromagnetic interference is great.
1 2 3
4
5
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...