The Basics
2-5
2.2
External Appearance
2.2.1
Front Panel
Figure 2-1 Front Panel
This monitor is designed to comply with the requirements of relative international
safety standards (IEC60601-1, EN60601-2-27 and EN60601-2-30) for medical
electrical equipment. This monitor has floating inputs and is protected against the
effects of defibrillation and electrosurgery. When proper electrodes are used and
applied according to the manufacturer instructions, the screen display will recover
within 10 seconds after defibrillation.
The alarm indicator of this monitor complies with the requirement of EN60825-1
A11 Class 1 for LED. The LED indicator varies its flash color and frequency to
indicate different alarm levels. For details, please refer to the section of
6.2.1 Visual
Alarms
.
WARNING
z
Move or lift the monitor by the handle only. Do not use the patient cable
or the power cord to move or lift the monitor. It might cause the monitor
to fall, which might damage the monitor or injure the patient.
Handle
Alarm indicator
Display
Control panel
Control knob
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...