SpO
2
Monitoring
12-28
If poor perfusion affects performance, consider using the Oxisensor R-15 sensor; it
obtains measurements from the nasal septal anterior ethmoid artery, an artery
supplied by the internal carotid. This sensor may obtain measurements when
peripheral perfusion is relatively poor. For low peripheral perfusion, consider using
the Nellcor RS-10 sensor, which is applied to the forehead or temple. These are sites
that may be spared during peripheral vasoconstriction.
12.4.5 SpO
2
Setup Menu
Selecting the SPO
2
label in the parameter window opens the following menu.
Figure 12-9 SpO
2
Setup Menu
You can perform the following settings in this menu.
ALM
SpO
2
alarm on/off status
ON: When a SpO
2
alarm occurs, the monitor gives alarm
indications and stores the alarm;
OFF: When a SpO
2
alarm occurs, the monitor neither
gives alarm indications nor stores the alarm;
When OFF is selected, the
icon is displayed on the
right of the SPO2 label.
ALM LEV
Alarm level
Options: HIGH, MED and LOW.
ALM REC
Alarm recording
ON: When a SpO
2
alarm occurs, the monitor enables the
recording;
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...