SpO
2
Monitoring
12-14
%SpO
2
100
97
95
90
85
80
75
70
65
DST
TM
Relative Correlation Canceler
Energy Output
Discrete Saturation Transform
The DST plot has two peaks: the peak corresponding to the higher saturation is
selected as the SpO
2
value. This entire sequence is repeated once every two seconds
on the most recent four seconds of raw data. The SpO
2
value therefore corresponds
to a running average of arterial hemoglobin saturation that is updated every two
seconds.
12.3.2 Precautions
WARNING
z
The pulse wave from the MS-7 module must NOT be used for apnea
monitoring.
z
As a trend towards patient deoxygenation is indicated, blood samples
should be analyzed by a laboratory co-oximeter to completely
understand the patient’s condition.
z
If an alarm condition (other than exceptions listed herein) occurs in the
Alarms Silenced status, the monitor only gives visual alarm symbols.
z
Measure the monitor’s leakage current whenever an external device is
connected to the serial port. Leakage current must not exceed 100
microamperes.
Содержание PM-9000
Страница 1: ...PM 9000 Patient Monitor Operation Manual...
Страница 2: ......
Страница 14: ...Contents XII FOR YOUR NOTES...
Страница 42: ...The Basics 2 18 FOR YOUR NOTES...
Страница 54: ...Installation and Maintenance 3 12 FOR YOUR NOTES...
Страница 106: ...Alarms 6 12 FOR YOUR NOTES...
Страница 111: ...8 1 8 Recording 8 1 Overview 8 2 8 2 Recording Types 8 2 8 3 Recorder Operations 8 5 8 4 Installing Recorder Paper 8 9...
Страница 131: ...10 1 10 Drug Calculation 10 1 Drug Calculation 10 2 10 2 Titration Table 10 5...
Страница 184: ...ECG RESP Monitoring 11 48 FOR YOUR NOTES...
Страница 216: ...SpO2 Monitoring 12 32 FOR YOUR NOTES...
Страница 276: ...CO Monitoring 16 14 FOR YOUR NOTES...
Страница 377: ......
Страница 378: ...P N 9100 20 11311 6 2...